Validation of extravascular lung water measurement by single transpulmonary thermodilution: human autopsy study
- PMID: 20819213
- PMCID: PMC3219254
- DOI: 10.1186/cc9250
Validation of extravascular lung water measurement by single transpulmonary thermodilution: human autopsy study
Abstract
Introduction: Gravimetric validation of single-indicator extravascular lung water (EVLW) and normal EVLW values has not been well studied in humans thus far. The aims of this study were (1) to validate the accuracy of EVLW measurement by single transpulmonary thermodilution with postmortem lung weight measurement in humans and (2) to define the statistically normal EVLW values.
Methods: We evaluated the correlation between pre-mortem EVLW value by single transpulmonary thermodilution and post-mortem lung weight from 30 consecutive autopsies completed within 48 hours following the final thermodilution measurement. A linear regression equation for the correlation was calculated. In order to clarify the normal lung weight value by statistical analysis, we conducted a literature search and obtained the normal reference ranges for post-mortem lung weight. These values were substituted into the equation for the correlation between EVLW and lung weight to estimate the normal EVLW values.
Results: EVLW determined using transpulmonary single thermodilution correlated closely with post-mortem lung weight (r = 0.904, P < 0.001). A linear regression equation was calculated: EVLW (mL) = 0.56 × lung weight (g) - 58.0. The normal EVLW values indexed by predicted body weight were approximately 7.4 ± 3.3 mL/kg (7.5 ± 3.3 mL/kg for males and 7.3 ± 3.3 mL/kg for females).
Conclusions: A definite correlation exists between EVLW measured by the single-indicator transpulmonary thermodilution technique and post-mortem lung weight in humans. The normal EVLW value is approximately 7.4 ± 3.3 mL/kg.
Trial registration: UMIN000002780.
Figures
Comment in
-
Calculation of the normal range of extravascular lung water.Crit Care. 2010;14(5):448; author reply 448. doi: 10.1186/cc9298. Epub 2010 Oct 29. Crit Care. 2010. PMID: 21067550 Free PMC article. No abstract available.
References
-
- Meade MO, Cook RJ, Guyatt GH, Groll R, Kachura JR, Bedard M, Cook DJ, Slutsky AS, Stewart TE. Interobserver variation in interpreting chest radiographs for the diagnosis of acute respiratory distress syndrome. Am J Respir Crit Care Med. 2000;161:85–90. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources