Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2010;14(5):R162.
doi: 10.1186/cc9250. Epub 2010 Sep 6.

Validation of extravascular lung water measurement by single transpulmonary thermodilution: human autopsy study

Affiliations
Comparative Study

Validation of extravascular lung water measurement by single transpulmonary thermodilution: human autopsy study

Takashi Tagami et al. Crit Care. 2010.

Abstract

Introduction: Gravimetric validation of single-indicator extravascular lung water (EVLW) and normal EVLW values has not been well studied in humans thus far. The aims of this study were (1) to validate the accuracy of EVLW measurement by single transpulmonary thermodilution with postmortem lung weight measurement in humans and (2) to define the statistically normal EVLW values.

Methods: We evaluated the correlation between pre-mortem EVLW value by single transpulmonary thermodilution and post-mortem lung weight from 30 consecutive autopsies completed within 48 hours following the final thermodilution measurement. A linear regression equation for the correlation was calculated. In order to clarify the normal lung weight value by statistical analysis, we conducted a literature search and obtained the normal reference ranges for post-mortem lung weight. These values were substituted into the equation for the correlation between EVLW and lung weight to estimate the normal EVLW values.

Results: EVLW determined using transpulmonary single thermodilution correlated closely with post-mortem lung weight (r = 0.904, P < 0.001). A linear regression equation was calculated: EVLW (mL) = 0.56 × lung weight (g) - 58.0. The normal EVLW values indexed by predicted body weight were approximately 7.4 ± 3.3 mL/kg (7.5 ± 3.3 mL/kg for males and 7.3 ± 3.3 mL/kg for females).

Conclusions: A definite correlation exists between EVLW measured by the single-indicator transpulmonary thermodilution technique and post-mortem lung weight in humans. The normal EVLW value is approximately 7.4 ± 3.3 mL/kg.

Trial registration: UMIN000002780.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Correlation of extravascular lung water (EVLW) measured by single transpulmonary thermodilution and by postmortem lung weight. EVLW (in milliliters) = [0.56 × lung weight (in grams)] - 58.0. n = 30, r = 0.90, P < 0.001. Line of identity is dashed.

Comment in

References

    1. Onwuanyi A, Taylor M. Acute decompensated heart failure: pathophysiology and treatment. Am J Cardiol. 2007;99:25D–30D. doi: 10.1016/j.amjcard.2006.12.017. - DOI - PubMed
    1. Atabai K, Matthay MA. The pulmonary physician in critical care. 5: Acute lung injury and the acute respiratory distress syndrome: definitions and epidemiology. Thorax. 2002;57:452–458. doi: 10.1136/thorax.57.5.452. - DOI - PMC - PubMed
    1. Ware LB, Matthay MA. Clinical practice. Acute pulmonary edema. N Engl J Med. 2005;353:2788–2796. doi: 10.1056/NEJMcp052699. - DOI - PubMed
    1. Rubenfeld GD, Caldwell E, Granton J, Hudson LD, Matthay MA. Interobserver variability in applying a radiographic definition for ARDS. Chest. 1999;116:1347–1353. doi: 10.1378/chest.116.5.1347. - DOI - PubMed
    1. Meade MO, Cook RJ, Guyatt GH, Groll R, Kachura JR, Bedard M, Cook DJ, Slutsky AS, Stewart TE. Interobserver variation in interpreting chest radiographs for the diagnosis of acute respiratory distress syndrome. Am J Respir Crit Care Med. 2000;161:85–90. - PubMed