Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2010 Jul;123(13):1645-51.

Echocardiographic mapping of left ventricular resynchronization during cardiac resynchronization therapy procedures

Affiliations
  • PMID: 20819622
Clinical Trial

Echocardiographic mapping of left ventricular resynchronization during cardiac resynchronization therapy procedures

Ngai-Yin Chan et al. Chin Med J (Engl). 2010 Jul.

Abstract

Background: Cardiac resynchronization therapy (CRT) is an effective electrical therapy for patients with moderate to severe heart failure and cardiac dyssynchrony. This study aimed to investigate the degree of acute left ventricular (LV) resynchronization with biventricular pacing (BVP) at different LV sites and to examine the feasibility of performing transthoracic echocardiography (TTE) to quantify acute LV resynchronization during CRT procedure.

Methods: Fourteen patients with NYHA Class III-IV heart failure, LV ejection fraction < or = 35%, QRS duration > or = 120 ms and septal-lateral delay (SLD) > or = 60 ms on tissue Doppler imaging (TDI), underwent CRT implant. TDI was obtained from three apical views during BVP at each accessible LV site and SLD during BVP was derived. Synchronicity gain index (Sg) by SLD was defined as (1 + (SLD at baseline--SLD at BVP)/SLD at baseline).

Results: Seventy-two sites were studied. Positive resynchronization (R+, Sg > 1) was found in 42 (58%) sites. R+ was more likely in posterior or lateral than anterior LV sites (66% vs. 36%, P < 0.001). Concordance of empirical LV lead implantation sites and sites with R+ was 50% (7/14).

Conclusions: The degree of acute LV resynchronization by BVP depends on LV lead location and empirical implantation of LV lead results in only 50% concordance with R+. Performing TTE during CRT implantation is feasible to identify LV sites with positive resynchronization.