. 2012 Jan;26(5):565-79.
doi: 10.1177/0885328210378057. Epub 2010 Sep 6.

Biomechanical characterization of an osteoporotic artificial bone model for the distal femur

Affiliations

Biomechanical characterization of an osteoporotic artificial bone model for the distal femur

Dirk Wähnert et al. J Biomater Appl. 2012 Jan.

Abstract

The treatment of osteoporotic distal femur fractures is still an unsolved problem of trauma surgery. The poor bone stock often leads to secondary loss of reduction and implant failure. Therefore, the development of new implants and their biomechanical testing is essential. In a previous study, we developed and initially characterized an artificial osteoporotic bone model of the distal femur. This follow-up study was performed to characterize this model in a biomechanical comparison. We investigated two different artificial bones: five foam cortical shell (Sawbones) and 10 custom-made artificial femoral condyles. Additionally, eight human femora were used for comparison. For biomechanical testing, two intramedullary nails (distal femur nail (DFN) and supracondylar nail (SCN)) were cyclically axial loaded in an AO 33 C2 unstable distal femoral fracture model. In our testing, the artificial bone showed a decrease in the axial stiffness of 27% for the SCN and 28% for the DFN compared to the human results. Also the number of cycles for a deformation of 2.5 mm was reduced by 55% (SCN) and 62% (DFN). This decrease was homogenous and caused by the relative high bone mineral density of the human specimen used. The modes of failure showed no difference between the artificial and human bones. Our customized artificial bone provides suitable results. In relation to the human bones classified as mildly osteoporotic, we assume that the biomechanical properties match to serve as an osteoporotic bone. Yet, we suggest to check transferability of the results with human material.