Variability in PET quantitation within a multicenter consortium

Affiliations

Variability in PET quantitation within a multicenter consortium

Frederic H Fahey et al. Med Phys. 2010 Jul.

Abstract

Purpose: The purpose of this study was to evaluate the variability in quantitation of positron emission tomography (PET) data acquired within the context of a multicenter consortium.

Methods: PET quantitation phantoms designed by American Association of Physicists in Medicine/ Society of Nuclear Medicine Task Group 145 were sent to the ten member sites of the Pediatric Brain Tumor Consortium (PBTC), a NIH-funded research consortium investigating the biology and therapies for brain tumors in children. The phantoms were water-filled cylinders (18.6 cm inside height and 20.4 cm inside diameter) based on the standard ACR phantom with four small, "hot" cylinders of varying diameters (8, 12, 16, 25 mm, all with 38 mm height), consisting of an equilibrium mixture of 68Ge/68Ga in an epoxy matrix. At each site, the operator added the appropriate amount of 18F to the water in the background in order to attain a feature-to-background ratio of roughly 4:1. The phantom was imaged and reconstructed as if it were a brain PET scan for the PBTC. An approximately 12 mm circular region of interest (ROI) was placed over each feature and in a central area in the background. The mean and maximum pixel values for each ROI were requested from local sites in units of activity concentration (Bq/ml) and the standard uptake value (SUV) (g/mL) based on bodyweight. The activity concentration was normalized by the decay-corrected known activity concentration for the features, and reported as the absolute recovery coefficient (RC). In addition, central analyses were performed by two observers

Results: The ten sites successfully imaged the phantom within 5 months and submitted the quantitative results and the phantom image data to the PBTC Operations and Biostatistics Center. The local site-based and central analyses yielded similar mean values for RC. Local site-based SUV measurements of the hot cylindrical features yielded greater variability than central analysis (COV range of 29.9%-42.8% compared to 7.7%-23.2%). Correcting for miscalculations in the local site reported SUVs substantially reduced the variation to levels similar to the central analysis (COV range of 8.8%-18.4%) and also led to the local sites providing a similar mean of the SUV values to those from the central analysis. In the central analysis, the use of mean SUV in place of maximum SUV for an ROI of fixed size substantially reduced the variation in the SUV values (COV ranges of 7.7%-11.3% vs. 9.3%-23.2%).

Conclusions: Based on this investigation, a SUV variability in the range of 10%-25% due solely to instrument and analysis factors can be expected in the context of a multicenter consortium if a central reading is used and quality assurance and quality control procedures are followed. The overall SUV variability can be expected to be larger than this due to biological and protocol factors.

PubMed Disclaimer

Figures

Figure 1
Figure 1
PET quantitation phantom. The phantom used in this investigation (top) is based on the phantom used for the ACR PET accreditation program, except that the hot cylindrical features were filled with 68Ge∕68Ga mixed with epoxy. At the time of imaging, the background portion of the phantom was filled with the appropriate activity concentration to produce an approximately 4:1 feature-to-background ratio. The ROI used for the analysis of these data was drawn to be the same size as the second smallest feature (bottom image designated by white arrow) which had a 1.2 cm diameter. Similar sized ROIs were placed over each feature as well one placed in the central background region.
Figure 2
Figure 2
Mean SUVs. The mean SUV within the ROI is plotted for Observers 1 and 2, as well as for the analysis done onsite as a function of cylindrical feature diameter.

References

    1. Boellaard R., Oyen W. J. G., Hoekstra C. J., Hoekstra O. S., Visser E. P., Willemsen A. T., Arends B., Verzijlbergen F. J., Zijlstra J., Paans A. M., Comans E. F. I., and Pruim J., “The Netherlands protocol for standardization and quantification of FDG whole body PET studies in multi-centre trials,” Eur. J. Nucl. Med. Mol. Imaging EJNMA6 35, 2320–2333 (2008).10.1007/s00259-008-0874-2 - DOI - PubMed
    1. Boellaard R., “Standards for PET image acquisition and quantitative data analysis,” J. Nucl. Med. JNMEAQ 50, 11S–20S (2009).10.2967/jnumed.108.057182 - DOI - PubMed
    1. Boellaard R. et al., Eur. J. Nucl. Med. Mol. Imaging EJNMA6 37, 181–200 (2010).10.1007/s00259-009-1297-4 - DOI - PMC - PubMed
    1. Jacene H. A., Leboulleux S., Baba S., Chatzifotiadis D., Goudarzi B., Teytelbaum O., Horton K. M., Kamel I., Macura K. J., Tsai H. L., Kowalski J., and Wahl R. L., “Assessment of interobserver reproducibility in quantitative 18F-FDG PET and CT measurements of tumor response to therapy,” J. Nucl. Med. JNMEAQ 50, 1760–1769 (2009).10.2967/jnumed.109.063321 - DOI - PubMed
    1. Young Poussaint T., Phillips P. C., Vajapeyam S., Fahey F. H., Robertson R. L., Osganian S., Ramamurthy U., Mulkern R. V., Treves S. T., Boyett J. M., and Kun L. E., “The neuroimaging center of the pediatric brain tumor consortium—Collaborative neuroimaging in pediatric brain tumor research: A works in progress,” AJNR Am. J. Neuroradiol. ZZZZZZ 28, 603–607 (2007). - PMC - PubMed

Publication types

LinkOut - more resources