In vivo performance of a laser fluorescence device for the approximal detection of caries in permanent molars

Affiliations
Randomized Controlled Trial

In vivo performance of a laser fluorescence device for the approximal detection of caries in permanent molars

K C Huth et al. J Dent. 2010 Dec.

Abstract

Objectives: The aim of this randomised clinical trial was to investigate if a laser fluorescence device is able to discriminate between sound and carious approximal sites and between enamel and dentinal lesions, as well as to find appropriate cut-off values.

Methods: One hundred and seventeen sound or uncavitated carious sites in permanent molars were visually and radiographically examined, then either opened or not, after which their laser fluorescence was measured. Forty-three lesions were opened, the caries removed and the clinically identified caries depths were registered in addition to the radiographical scoring. Seventy-four sites were radiographically deemed sound or had enamel caries and were not opened. Here, the radiographical scorings were registered.

Results: Taking the radiographic scoring as gold standard for all investigated approximal sites, sound sites (D(0), n=40) showed significantly lower laser fluorescence measurements than carious sites (D(1-4), n=77) (Mann-Whitney test, P<0.025) suggesting a cut-off at 7 (sensitivity=0.68, specificity=0.7). Comparing measurements of D(0-2) (n=74) and D(3,4) (n=43), the results were also different by a statistically significant amount (P<0.025) and the cut-off calculated to be 16 (sensitivity=0.6, specificity=0.84). A fair positive correlation between laser fluorescence values and radiographical scoring was found (rho=+0.47, P<0.01). Analysing the 43 opened lesions with their clinically found lesion depths as gold standard, there was a fair positive correlation to the laser fluorescence values (rho=+0.34, P=0.03) and a moderately strong correlation to the radiographic scoring (rho=+0.67, P<0.01).

Conclusion: The device may be an adjunct tool in the approximal detection of caries along with established procedures.