The revised digital transcutaneous PCO2/SpO2 ear sensor is a reliable noninvasive monitoring tool in patients after cardiac surgery
- PMID: 20851636
- DOI: 10.1053/j.jvca.2010.06.021
The revised digital transcutaneous PCO2/SpO2 ear sensor is a reliable noninvasive monitoring tool in patients after cardiac surgery
Abstract
Objective: The aim of this study was to validate the revised SenTec V-Sign 2 sensor (SenTec AG, Therwil, Switzerland) for combined noninvasive continuous assessment of pulse rate, pulse oximetry (SpO(2)), and transcutaneous carbon dioxide tension (PtcCO(2)) in adults after cardiac surgery.
Design: A prospective clinical study.
Setting: A single-center university hospital.
Participants: Twenty adult patients aged 36 to 84 years after cardiac surgery.
Interventions: SpO(2) and PtcCO(2) values of three V-Sign 2 sensors (SenTec AG) attached at the earlobe, forehead, and cheek and SpO(2) values of the Nellcor Durasensor (Model DS-100A; Nellcor Puritan Bennett Inc, Pleasanton, CA) were compared with simultaneous measurements of blood gases and end-expiratory carbon dioxide.
Measurements and main results: Measurements were performed during periods of hyper-, normo-, and hypocapnia and then at 30-minute intervals up to 5 hours. Bland-Altman analysis and simple regression analysis were used.
Results: The detection failures for PtcCO(2) were 0.3% to 1.3%, for SpO(2) 10% to 25%, and for pulse rate 5% to 10%. The V-Sign 2 earlobe sensor provided the best results. The mean bias and limits of agreement for PtcCO(2ear) and PaCO(2) were 1.1 and -3.4/+5.5 mmHg. The drift of PtcCO(2) was negligible at all locations. The mean bias and limits of agreement of V-Sign SpO(2ear) and SaO(2), as well as V-Sign pulse rate and the electrocardiogram, were -1.7% and -6.8/+3.9% and 1.2 beats/min and -3.3/+5.8 beats/min. End-expiratory carbon dioxide showed a weak correlation with PaCO(2) (r(2) = 0.47).
Conclusions: Transcutaneous capnometry using the revised V-Sign 2 sensor at the earlobe is a reliable monitoring tool during the recovery period of patients after cardiac surgery. This approach has the potential to reduce the number of arterial blood gas samples.
Copyright © 2011 Elsevier Inc. All rights reserved.
Similar articles
-
Clinical validation of a digital transcutaneous PCO2/SpO2 ear sensor in adult patients after cardiac surgery.J Clin Monit Comput. 2007 Oct;21(5):303-9. doi: 10.1007/s10877-007-9088-6. Epub 2007 Aug 16. J Clin Monit Comput. 2007. PMID: 17701383
-
Evaluation of a new combined transcutaneous measurement of PCO2/pulse oximetry oxygen saturation ear sensor in newborn patients.Pediatrics. 2005 Jan;115(1):e64-8. doi: 10.1542/peds.2004-0946. Epub 2004 Dec 15. Pediatrics. 2005. PMID: 15601814
-
An evaluation of a new combined Spo2/PtcCO2 sensor in very low birth weight infants.Anesth Analg. 2008 Jul;107(1):125-9. doi: 10.1213/ane.0b013e3181733e47. Anesth Analg. 2008. PMID: 18635477
-
Monitoring carbon dioxide in critical care: the newest vital sign?Crit Care Nurs Clin North Am. 2004 Sep;16(3):445-51, x. doi: 10.1016/j.ccell.2004.05.002. Crit Care Nurs Clin North Am. 2004. PMID: 15358391 Review.
-
Accuracy and precision of transcutaneous carbon dioxide monitoring: a systematic review and meta-analysis.Thorax. 2019 Feb;74(2):157-163. doi: 10.1136/thoraxjnl-2017-211466. Epub 2018 Sep 12. Thorax. 2019. PMID: 30209079
Cited by
-
Carbon dioxide insufflation in colonoscopy is safe: a prospective trial of 347 patients.Diagn Ther Endosc. 2012;2012:692532. doi: 10.1155/2012/692532. Epub 2012 Sep 29. Diagn Ther Endosc. 2012. PMID: 23055655 Free PMC article.
-
Noninvasive Measurement of Carbon Dioxide during One-Lung Ventilation with Low Tidal Volume for Two Hours: End-Tidal versus Transcutaneous Techniques.PLoS One. 2015 Oct 14;10(10):e0138912. doi: 10.1371/journal.pone.0138912. eCollection 2015. PLoS One. 2015. PMID: 26466140 Free PMC article.
-
Transcutaneous continuous carbon dioxide tension monitoring reduced incidence, degree and duration of hypercapnia during combined regional anaesthesia and monitored anaesthesia care in shoulder surgery patients.J Clin Monit Comput. 2015 Aug;29(4):499-507. doi: 10.1007/s10877-014-9627-x. Epub 2014 Oct 14. J Clin Monit Comput. 2015. PMID: 25312782 Clinical Trial.
-
Transcutaneous carbon dioxide monitoring to avoid hypercapnia during complex catheter ablations: a feasibility study.J Interv Card Electrophysiol. 2015 Sep;43(3):307-11. doi: 10.1007/s10840-015-0013-5. Epub 2015 May 22. J Interv Card Electrophysiol. 2015. PMID: 25997689
-
Accuracy of Transcutaneous Carbon Dioxide Measurement During Transcatheter Aortic Valve Replacement Under Monitored Anesthesia Care: A Prospective Observational Study.Cureus. 2024 Feb 5;16(2):e53661. doi: 10.7759/cureus.53661. eCollection 2024 Feb. Cureus. 2024. PMID: 38455779 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical