Randomized Controlled Trial
. 2010 Dec;3(6):571-7.
doi: 10.1161/CIRCEP.110.936161. Epub 2010 Sep 18.

A prospective comparison of cardiac imaging using intracardiac echocardiography with transesophageal echocardiography in patients with atrial fibrillation: the intracardiac echocardiography guided cardioversion helps interventional procedures study

Affiliations
Randomized Controlled Trial

A prospective comparison of cardiac imaging using intracardiac echocardiography with transesophageal echocardiography in patients with atrial fibrillation: the intracardiac echocardiography guided cardioversion helps interventional procedures study

Sanjeev Saksena et al. Circ Arrhythm Electrophysiol. 2010 Dec.

Abstract

Background: The Intracardiac Echocardiography Guided Cardioversion Helps Interventional Procedures study evaluated the concordance of intracardiac echocardiography (ICE) with transesophageal echocardiography (TEE) in patients with atrial fibrillation (AF).

Methods and results: Patients with AF undergoing right heart catheterization underwent left atrium (LA) and interatrial septal (IAS) imaging by TEE and ICE. A blinded comparison of the 2 modalities was performed at a core laboratory. Ninety-five patients aged 58 ± 12 years completed the study. The LA was profiled in all patients with both techniques, and concordance for image quality was 96%. LA appendage (LAA) imaging was achieved in 85% with ICE and 96% with TEE. There was no difference in the presence of spontaneous echo contrast between ICE and TEE during LA imaging, but there was a trend toward a greater incidence in the LAA with TEE (P = 0.109). Intracardiac thrombus was uncommonly seen (TEE, 6.9%; ICE, 5.2%). The concordance for the presence or absence of thrombus was 97% in the LA and 92% in the LAA, but the latter was detected more frequently with TEE. IAS imaging was achieved in 91% with ICE and in 97% with TEE (P = 0.177). Concordance for patent foramen ovale and atrial septal aneurysms was 100% and 96%, respectively. A negative ICE examination was associated with absence of dense echo contrast or thrombus on TEE in 86%.

Conclusions: This study provides validation for the use of ICE for LA and IAS imaging. ICE imaging was less sensitive compared to TEE for LAA thrombus identification. Clinical Trial Registration- URL: http://www.clinicaltrials.gov. Unique identifier: NCT00281073.