. 2011 Apr;21(4):882-9.
doi: 10.1007/s00330-010-1957-6. Epub 2010 Sep 21.

Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT

Affiliations

Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT

Tobias Struffert et al. Eur Radiol. 2011 Apr.

Abstract

Objective: We tested the hypothesis that Flat Detector computed tomography (FD-CT) with intravenous contrast medium would allow the calculation of whole brain cerebral blood volume (CBV) mapping (FD-CBV) and would correlate with multislice Perfusion CT (PCT).

Methods: Twenty five patients were investigated with FD-CBV and PCT. Correlation of the CBV maps of both techniques was carried out with measurements from six anatomical regions from both sides of the brain. Mean values of each region and the correlation coefficient were calculated. Bland-Altman analysis was performed to compare the two different imaging techniques.

Results: The image and data quality of both PCT and FD-CBV were suitable for evaluation in all patients. The mean CBV values of FD-CBV and PCT showed only minimal differences with overlapping standard deviation. The correlation coefficient was 0.79 (p < 0.01). Bland-Altman analysis showed a mean difference of -0.077 ± 0.48 ml/100 g between FD-CBV and PCT CBV measurements, indicating that FD-CBV values were only slightly lower than those of PCT.

Conclusion: CBV mapping with intravenous contrast medium using Flat Detector CT compared favourably with multislice PCT. The ability to assess cerebral perfusion within the angiographic suite may improve the management of ischaemic stroke and evaluation of the efficacy of dedicated therapies.

PubMed Disclaimer

References

    1. Radiology. 2008 Jun;247(3):818-25 - PubMed
    1. Radiology. 2003 Jun;227(3):725-30 - PubMed
    1. Stroke. 2006 Apr;37(4):979-85 - PubMed
    1. Diagn Interv Radiol. 2005 Mar;11(1):10-3 - PubMed
    1. Eur J Radiol. 1999 Jun;30(3):170-84 - PubMed