Prospective study of access site complications of automated contrast injection with peripheral venous access in MDCT
- PMID: 20858804
- DOI: 10.2214/AJR.09.3739
Prospective study of access site complications of automated contrast injection with peripheral venous access in MDCT
Abstract
Objective: The purpose of this article is to prospectively assess the frequency and type of IV injection site complications associated with high-flow power injection of nonionic contrast medium in MDCT.
Subjects and methods: Contrast-enhanced (300-370 mg iodine/mL) MDCT examinations with high flow rates (up to 8 mL/s) using automatic CT injectors were performed according to standardized MDCT protocols. The location, type, and size (16-24 gauge) of IV catheters and volumes, iodine concentration, and flow rates of contrast medium were documented. Patients were questioned about associated discomfort, IV catheter sites were checked, and adverse effects were recorded.
Results: Prospectively, 4,457 patients were studied. The injection rate ranged from 1-2.9 mL/s (group 1; n = 1,140) to 3-4.9 mL/s (group 2; n = 2,536) to 5-8 mL/s (group 3; n = 781); 1.2% of the patients experienced extravasations (n = 52). Contrast medium iodine concentration, flow rates, and volumes were not related to the frequency of extravasation. The extravasation rate was highest with 22-gauge IV catheters (2.2%; p < 0.05) independently of the anatomic location. For 20-gauge IV catheters, extravasation rates were significantly higher in the dorsum of the hand than in the antecubital fossa (1.8% vs 0.8%; p = 0.018). Extravasation rates were higher in older patients (≥ 50 vs < 50 years, 0.6% vs 1.4%; p = 0.019). Different iodine concentrations did not trigger significant differences in contrast material reactions (p = 0.782).
Conclusion: Automated IV contrast injection applying high flow rates (i.e., up to 8 mL/s) is performed without increased risk of extravasation. The overall extravasation rate was 1.2% and showed no correlation with iodine concentration, flow rates, or contrast material reactions. Performing high flow rates with low-diameter IV catheters (e.g., 22-gauge catheters) and a location of IV catheter in the hand is associated with a higher extravasation rate.
Similar articles
-
Peripheral intravenous power injection of iodinated contrast media through 22G and 20G cannulas: can high flow rates be achieved safely? A clinical feasibility study.Rofo. 2009 Apr;181(4):355-61. doi: 10.1055/s-0028-1109181. Epub 2009 Mar 11. Rofo. 2009. PMID: 19280548 Clinical Trial.
-
Safety of power injector use in children as measured by incidence of extravasation.AJR Am J Roentgenol. 2006 Aug;187(2):580-3. doi: 10.2214/AJR.05.0667. AJR Am J Roentgenol. 2006. PMID: 16861567 Clinical Trial.
-
Contrast media reactions and extravasation: relationship to intravenous injection rates.Radiology. 1998 Nov;209(2):411-6. doi: 10.1148/radiology.209.2.9807567. Radiology. 1998. PMID: 9807567
-
CT contrast extravasation in children: a single-center experience and systematic review.Pediatr Radiol. 2024 Jan;54(1):34-42. doi: 10.1007/s00247-023-05811-6. Epub 2023 Nov 22. Pediatr Radiol. 2024. PMID: 37991500
-
Risk of extravasation after power injection of contrast media via the proximal port of multilumen central venous catheters: case report and review of the literature.Rofo. 2010 Jan;182(1):14-9. doi: 10.1055/s-0028-1109742. Epub 2009 Oct 26. Rofo. 2010. PMID: 19859861 Review.
Cited by
-
Extravasation of iodinated contrast medium in cancer patients undergoing computed tomography.Radiol Bras. 2018 Jul-Aug;51(4):236-241. doi: 10.1590/0100-3984.2017.0064. Radiol Bras. 2018. PMID: 30202127 Free PMC article.
-
More holes, more contrast? Comparing an 18-gauge non-fenestrated catheter with a 22-gauge fenestrated catheter for cardiac CT.PLoS One. 2020 Jun 8;15(6):e0234311. doi: 10.1371/journal.pone.0234311. eCollection 2020. PLoS One. 2020. PMID: 32511272 Free PMC article.
-
Two Small Intravenous Catheters for High-Rate Contrast Medium Injection for Computed Tomography in Patients Lacking Superficial Veins to Accommodate a Large Catheter.Korean J Radiol. 2018 May-Jun;19(3):489-497. doi: 10.3348/kjr.2018.19.3.489. Epub 2018 Apr 6. Korean J Radiol. 2018. PMID: 29713227 Free PMC article.
-
Clinical pharmacology, uses, and adverse reactions of iodinated contrast agents: a primer for the non-radiologist.Mayo Clin Proc. 2012 Apr;87(4):390-402. doi: 10.1016/j.mayocp.2012.01.012. Mayo Clin Proc. 2012. PMID: 22469351 Free PMC article. Review.
-
We Built This House; It's Time to Move in: Leveraging Existing DICOM Structure to More Completely Utilize Readily Available Detailed Contrast Administration Information.J Digit Imaging. 2015 Aug;28(4):407-11. doi: 10.1007/s10278-015-9771-y. J Digit Imaging. 2015. PMID: 25700615 Free PMC article. Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical