Comparative Study
. 2010 Nov;17(11):1734-8.
doi: 10.1128/CVI.00332-10. Epub 2010 Sep 22.

Multiplex detection of IgM and IgG class antibodies to Toxoplasma gondii, rubella virus, and cytomegalovirus using a novel multiplex flow immunoassay

Affiliations
Comparative Study

Multiplex detection of IgM and IgG class antibodies to Toxoplasma gondii, rubella virus, and cytomegalovirus using a novel multiplex flow immunoassay

M J Binnicker et al. Clin Vaccine Immunol. 2010 Nov.

Abstract

The goal of this study was to evaluate the BioPlex 2200 Toxoplasma, rubella, and cytomegalovirus (CMV) (ToRC) IgG and IgM multiplex immunoassays (Bio-Rad Laboratories, Hercules, CA) and compare the results to those of conventional testing by enzyme immunoassay (EIA) and enzyme-linked fluorescent assay (ELFA). Serum specimens (n = 600) submitted for routine ToRC IgG and IgM testing by EIA (SeraQuest, Doral, FL; Diamedix, Miami, FL) or ELFA (Vidas; bioMérieux, Durham, NC) were also tested by the BioPlex ToRC multiplex immunoassays. Samples showing discordant results were retested by both methods, with further discrepancies being arbitrated by a third assay. Following repeat testing, the BioPlex Toxoplasma, rubella, and CMV IgG assays demonstrated agreements of 98.7 (592/600 specimens), 93.3 (560/600 specimens), and 98.3% (590/600 specimens), respectively, while the ToRC IgM assays yielded agreements of 91.2 (547/600 specimens), 87.3 (524/600 specimens), and 95.2% (571/600 specimens), respectively. The BioPlex ToRC IgG assays provided results comparable to EIA/ELFA results, with kappa coefficients showing near-perfect agreement for the Toxoplasma (κ = 0.94) and CMV (κ = 0.97) IgG assays and substantial agreement for the rubella IgG assay (κ = 0.66). The BioPlex ToRC IgM assays showed lower specificity with only slight agreement for Toxoplasma IgM (κ = 0.07), poor agreement for rubella IgM (κ = -0.03), and moderate agreement for CMV IgM (κ = 0.55). Both the BioPlex IgG and IgM assays reduced turnaround time (1.7 h versus 5.5 h by EIA/ELFA for 100 specimens) and eliminated the necessity to manually pipette or aliquot specimens prior to testing.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Abdel-Fattah, S. A., A. Bhat, S. Illanes, J. L. Bartha, and D. Carrington. 2005. TORCH test for fetal medicine indications: only CMV is necessary in the United Kingdom. Prenat. Diagn. 25:1028-1031. - PubMed
    1. Averhoff, F., J. Zucker, C. Vellozzi, S. Redd, C. Woodfill, S. Waterman, J. Baggs, M. Weinberg, A. Rodriquez-Lainz, V. Carrion, C. Goto, and S. E. Reef. 2006. Adequacy of surveillance to detect endemic rubella transmission in the United States. Clin. Infect. Dis. 43(Suppl. 3):S151-S157. - PubMed
    1. Berth, M., and E. Bosmans. 2010. Comparison of three automated immunoassay methods for the determination of Epstein-Barr virus-specific immunoglobulin M. Clin. Vaccine Immunol. 17:559-563. - PMC - PubMed
    1. Binnicker, M. J., D. J. Jespersen, and J. A. Harring. 2010. Evaluation of three multiplex flow immunoassays compared to an enzyme immunoassay for the detection and differentiation of IgG class antibodies to herpes simplex virus types 1 and 2. Clin. Vaccine Immunol. 17:253-257. - PMC - PubMed
    1. Doern, G. V., L. Robbie, and L. Marrama. 1994. Comparison of two enzyme immunoassays and two latex agglutination assays for detection of cytomegalovirus antibody. Diagn. Microbiol. Infect. Dis. 20:109-112. - PubMed