Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation

Affiliations
Comparative Study

Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation

Aissam Lyazidi et al. Intensive Care Med. 2010 Dec.

Abstract

Purpose: During volume-controlled ventilation, part of the volume delivered is compressed into the circuit. To correct for this phenomenon, modern ventilators use compensation algorithms. Humidity and temperature also influence the delivered volume.

Methods: In a bench study at a research laboratory in a university hospital, we compared nine ICU ventilators equipped with compensation algorithms, one with a proximal pneumotachograph and one without compensation. Each ventilator was evaluated under normal, obstructive, and restrictive conditions of respiratory mechanics. For each condition, three tidal volumes (V (T)) were set (300, 500, and 800 ml), with and without an inspiratory pause. The insufflated volume and the volume delivered at the Y-piece were measured independently, without a humidification device, under ambient temperature and pressure and dry gas conditions. We computed the actually delivered V (T) to the lung under body temperature and pressure and saturated water vapour conditions (BTPS).

Results: For target V (T) values of 300, 500, and 800 ml, actually delivered V (T) under BTPS conditions ranged from 261 to 396 ml (-13 to +32%), from 437 to 622 ml (-13 to +24%), and from 681 to 953 ml (-15 to +19%), respectively (p < 0.01). Respiratory system mechanics and application of an inspiratory pause significantly affected actually delivered V (T). Assuming a set V (T) of 6 ml/kg of predicted body weight, a difference of 1-2 ml/kg with actually delivered V (T) would be commonly observed.

Conclusion: The difference between preset V (T) and actually delivered V (T) is clinically meaningful and differs across modern ICU ventilators.

PubMed Disclaimer

References

    1. Am J Respir Crit Care Med. 1998 Dec;158(6):1831-8 - PubMed
    1. Anesthesiology. 1983 Nov;59(5):442-6 - PubMed
    1. Am J Respir Crit Care Med. 2004 Nov 15;170(10):1073-9 - PubMed
    1. Intensive Care Med. 2008 Sep;34(9):1669-75 - PubMed
    1. Am J Respir Crit Care Med. 2000 Dec;162(6):2109-12 - PubMed

LinkOut - more resources