Convex optimization of coincidence time resolution for a high-resolution PET system
- PMID: 20876008
- PMCID: PMC3693485
- DOI: 10.1109/TMI.2010.2080282
Convex optimization of coincidence time resolution for a high-resolution PET system
Abstract
We are developing a dual panel breast-dedicated positron emission tomography (PET) system using LSO scintillators coupled to position sensitive avalanche photodiodes (PSAPD). The charge output is amplified and read using NOVA RENA-3 ASICs. This paper shows that the coincidence timing resolution of the RENA-3 ASIC can be improved using certain list-mode calibrations. We treat the calibration problem as a convex optimization problem and use the RENA-3's analog-based timing system to correct the measured data for time dispersion effects from correlated noise, PSAPD signal delays and varying signal amplitudes. The direct solution to the optimization problem involves a matrix inversion that grows order (n(3)) with the number of parameters. An iterative method using single-coordinate descent to approximate the inversion grows order (n). The inversion does not need to run to convergence, since any gains at high iteration number will be low compared to noise amplification. The system calibration method is demonstrated with measured pulser data as well as with two LSO-PSAPD detectors in electronic coincidence. After applying the algorithm, the 511 keV photopeak paired coincidence time resolution from the LSO-PSAPD detectors under study improved by 57%, from the raw value of 16.3 ±0.07 ns full-width at half-maximum (FWHM) to 6.92 ±0.02 ns FWHM ( 11.52 ±0.05 ns to 4.89 ±0.02 ns for unpaired photons).
Figures
References
-
- Olcott PD, Lau FWY, Levin CS. Data acquisition system design for a 1 mm3 resolution PSAPD-based pet system, 2007 IEEE Nucl. Sci. Symp. Conf. Rec. 2007:3206–3211.
-
- Levin CS. New imaging technologies to enhance the molecular sensitivity of positron emission tomography. Proc. IEEE. 2008 Mar;96(no. 3):439–467.
-
- Tumer TO, Cajipe VB, Clajus M, Hayakawa S, Volkovskii a. Multi-channel front-end readout IC for position sensitive solid-state detectors. 2006 IEEE Nucl. Sci. Symp. Conf. Rec.; Oct. 29–Nov.4 2006; San Diego, CA.
-
- Li w, Du y, Yanoff BD, Gordon JS. Energy resolution limiting factors of multi-pixel events in 3D position sensitive CZT gamma-ray spectrometer, 2008 IEEE Nucl. Sci. Symp. Conf. Rec. 2008:496–502.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Miscellaneous
