Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2004 Jun;51(3):631-635.
doi: 10.1109/TNS.2004.830975.

FastSPECT II: A Second-Generation High-Resolution Dynamic SPECT Imager

Affiliations

FastSPECT II: A Second-Generation High-Resolution Dynamic SPECT Imager

Lars R Furenlid et al. IEEE Trans Nucl Sci. 2004 Jun.

Abstract

FastSPECT II is a recently commissioned 16-camera small-animal SPECT imager built with modular scintillation cameras and list-mode data-acquisition electronics. The instrument is housed in a lead-shielded enclosure and has exchangeable aperture assemblies and adjustable camera positions for selection of magnification, pinhole size, and field of view. The calibration of individual cameras and measurement of an overall system imaging matrix (1 mm(3) voxels) are supported via a five-axis motion-control system.Details of the system integration and results of characterization and performance measurements are presented along with first tomographic images. The dynamic imaging capabilities of the instrument are explored and discussed.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Assembly drawing and photograph of FastSPECT II 3 × 3 modular gamma-ray camera.
Fig. 2
Fig. 2
The list-mode data-acquisition components for a single FastSPECT II modular gamma-ray camera.
Fig. 3
Fig. 3
Photographs of FastSPECT II (a) prior to installation of shielding and (b) in finished condition together with calibration and positioning stage.
Fig. 4
Fig. 4
Measured mean response of the center PMT as a well collimated source is scanned across the face of one of the FSII cameras.
Fig. 5
Fig. 5
Measured MDRF for the three PMTs immediately underneath a horizontal scan of a FASTSPECT II modular gamma-ray camera close to the mid-line.
Fig. 6
Fig. 6
Phantom employed for first tomographic images along with a coin for scale. Total activity was ~1 mCi.
Fig. 7
Fig. 7
Tomographic slices of the phantom shown in Fig. 6. Each slice is 1 mm thick, and has lateral dimensions of 20 mm × 20 mm. Two spheres are at top edge of calibration volume.

References

    1. Klein WP, Barrett HH, Pang IW, Patton DD, Rogulski MM, Sain JD, Smith WE. FASTSPECT: Electrical and mechanical design of a high-resolution dynamic SPECT imager. Proc Conf Rec 1995 IEEE NSS/MIC. 1995;2:931–933.
    1. Kastis GK, Barber HB, Barrett HH, Gifford HC, Pang IW, Patton DD, Sain JD, Stevenson G, Wilson DW. High resolution SPECT imager for three-dimensional imaging of small animals. J Nucl Med. 1998;39(5-suppl):9–9. - PubMed
    1. Rogers JG, Saylor DP, Harrop R, Yao XG, Leitao CV, Pate BD. Design of an efficient position sensitive gamma ray detector for nuclear medicine. Phys Med Biol. 1986;31(10):1061–1090. - PubMed
    1. Milster TD, Selberg LA, Barrett HH, Easton RL, Rossi GR, Arendt J, Simpson RG. A modular scintillation camera for use in nuclear medicine. IEEE Trans Nucl Sci. 1984;NS-31:575–580.
    1. Milster TD, Aarsvold JN, Barrett HH, Landesman AL, Mar LS, Patton DD, Roney TJ, Rowe RK, Seacat RH., III A full-field modular gamma camera. J Nucl Med. 1990;5(31):632–639. - PubMed

LinkOut - more resources