. 2011 Jun;22(6):1907-13.
doi: 10.1007/s00198-010-1413-0. Epub 2010 Sep 29.

The effect of hierarchical micro/nanosurface titanium implant on osseointegration in ovariectomized sheep

Affiliations

The effect of hierarchical micro/nanosurface titanium implant on osseointegration in ovariectomized sheep

J Xiao et al. Osteoporos Int. 2011 Jun.

Abstract

Hydrofluoric etching and anodized hierarchical micro/nanotextured surface titanium implant was placed in mandibles of ovariectomized sheep for 12 weeks, and it showed improved osseointegration by resonance frequency analysis (RFA), microcomputed tomography (micro-CT) evaluation, histomorphometry, and biomechanical test.

Introduction: This study aimed to investigate the effects of micro/nanotextured titanium implant on osseointegration in ovariectomized (OVX) sheep.

Methods: The hierarchical micro/nanotextured surface of titanium implant was fabricated by acid in 0.5% (w/v) hydrofluoric (HF) and anodized in HF acid electrolytes with a DC power of 20 V, and the machined surface implants with no treatment served as control group. The implants were placed in mandibles of OVX sheep, respectively. Twelve weeks after implantation, RFA, microcomputed tomography, histomorphometry, and biomechanical tests were applied to detect the osseointegration of the two groups.

Results: The implant stability quotient (ISQ) values, the maximum pull-out forces, and the bone-implant contact (BIC) were 65.5 ± 6.3, 490.6 ± 72.7 N, and 58.31 ± 5.79% in the micro/nanogroup and 58.3 ± 8.9, 394.5 ± 54.5 N, and 46.85 ± 5.04% in the control group, respectively. There was no significant difference between the two groups in ISQ values (p > 0.05), but in the micro/nanogroup, the maximal pull-out force and the BIC were increased significantly (p < 0.05 or p < 0.01). Micro-CT analysis showed that the bone volume ratio and the trabecular number increased significantly (p < 0.01), and the trabecular separation decreased significantly (p < 0.05) in the micro/nanogroup.

Conclusion: Implant modification by HF acid etching and anodization to form a hierarchical micro/nanotextured surface could improve titanium implant osseointegration in OVX sheep 12 weeks after implantation.

PubMed Disclaimer

References

    1. J Prosthodont. 2009 Jun;18(4):309-23 - PubMed
    1. J Biomed Mater Res B Appl Biomater. 2008 Jul;86(1):162-9 - PubMed
    1. J Clin Periodontol. 2007 Jul;34(7):610-7 - PubMed
    1. Implant Dent. 2004 Mar;13(1):79-87 - PubMed
    1. Int J Prosthodont. 2003 Nov-Dec;16(6):602-8 - PubMed

Publication types

LinkOut - more resources