Dosimetric characteristics of a single use MOSFET dosimeter for in vivo dosimetry in proton therapy
- PMID: 20879587
- DOI: 10.1118/1.3467753
Dosimetric characteristics of a single use MOSFET dosimeter for in vivo dosimetry in proton therapy
Abstract
Purpose: Entrance dose (or skin dose) is an important part of patient quality assurance in external beam radiation therapy. However, entrance dose verification in proton beam is not routinely performed. In this study, the OneDose single use MOSFET detector system for in vivo dosimetry measurement in proton therapy is investigated.
Methods: Using a solid water phantom, several fundamental dosimetric characteristics of the OneDose system are studied with a proton beam: The reproducibility (consistency) of the dosimeter, the linearity with dose and dose rate, energy dependence, directional dependence, LET dependence, and fading (delay readout with time) is studied.
Results: OneDose detectors show dose and dose rate linearity but exhibit pronounced energy dependence at depth and a large variation in dose response with LET. On the other hand, the detector response remain relatively constant (within 3%) at surface over a wide range of energies. There is also a slight angular dependence (about 2%) up to 60 degrees angle of incidence. However, detector orientation such that incidence along the long axis of the detector should be avoided as the proton beam will have to traverse a large amount of the copper backing. Since most in vivo dosimetry involves entrance dose measurement, the OneDose at surface appears to be well suited for such application. OneDose exhibits small intrabatch variation (< or = 2% at one SD) indicating that it is only necessary to calibration a few detectors from each batch. The interbatch variation is generally within 3%.
Conclusions: The small detector size and its relatively flexible design of OneDose allow dose measurement to be performed on a curved surface or in small cavities that is otherwise difficult with the conventional diode detectors. The slight drawback in its angular dependence can be easily handled by angular dependence table. However, since OneDose is a single use detector, the intra-batch consistency must be verified before the remaining detectors from the same batch could be used for in vivo dosimetry. It is advisable that the detectors from the same batch be taken for the same application to reduce the dosimetric uncertainty. For detectors from different batches, inter-batch consistency should also be verified to obtain reliable results. OneDose provides an opportunity to measure in vivo dose with proton beam within acceptable clinical criterion of +/- (5.0%-6.5%).
Similar articles
-
Characterization of a cable-free system based on p-type MOSFET detectors for "in vivo" entrance skin dose measurements in interventional radiology.Med Phys. 2012 Aug;39(8):4866-74. doi: 10.1118/1.4736806. Med Phys. 2012. PMID: 22894413
-
Application of commercial MOSFET detectors for in vivo dosimetry in the therapeutic x-ray range from 80 kV to 250 kV.Phys Med Biol. 2005 Jan 21;50(2):289-303. doi: 10.1088/0031-9155/50/2/008. Phys Med Biol. 2005. PMID: 15742945
-
Commissioning and implementation of an implantable dosimeter for radiation therapy.J Appl Clin Med Phys. 2013 Mar 4;14(2):3989. doi: 10.1120/jacmp.v14i2.3989. J Appl Clin Med Phys. 2013. PMID: 23470929 Free PMC article.
-
A literature review of electronic portal imaging for radiotherapy dosimetry.Radiother Oncol. 2008 Sep;88(3):289-309. doi: 10.1016/j.radonc.2008.07.008. Epub 2008 Aug 14. Radiother Oncol. 2008. PMID: 18706727 Review.
-
In vivo dosimetry during external photon beam radiotherapy.Int J Radiat Oncol Biol Phys. 1999 Jan 15;43(2):245-59. doi: 10.1016/s0360-3016(98)00341-1. Int J Radiat Oncol Biol Phys. 1999. PMID: 10030247 Review.
Cited by
-
In vivo proton dosimetry using a MOSFET detector in an anthropomorphic phantom with tissue inhomogeneity.J Appl Clin Med Phys. 2012 Mar 8;13(2):3699. doi: 10.1120/jacmp.v13i2.3699. J Appl Clin Med Phys. 2012. PMID: 22402385 Free PMC article.
-
Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.J Med Phys. 2014 Jul;39(3):142-9. doi: 10.4103/0971-6203.139002. J Med Phys. 2014. PMID: 25190992 Free PMC article.
-
Proton dose distribution measurements using a MOSFET detector with a simple dose-weighted correction method for LET effects.J Appl Clin Med Phys. 2011 Apr 4;12(2):3431. doi: 10.1120/jacmp.v12i2.3431. J Appl Clin Med Phys. 2011. PMID: 21587191 Free PMC article.
-
General Purpose Transistor Characterized as Dosimetry Sensor of Proton Beams.Sensors (Basel). 2023 Apr 6;23(7):3771. doi: 10.3390/s23073771. Sensors (Basel). 2023. PMID: 37050831 Free PMC article.
-
Passively scattered proton beam entrance dosimetry with a plastic scintillation detector.Phys Med Biol. 2015 Feb 7;60(3):1185-98. doi: 10.1088/0031-9155/60/3/1185. Epub 2015 Jan 15. Phys Med Biol. 2015. PMID: 25591037 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources