Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses
- PMID: 20888199
- DOI: 10.1016/j.zemedi.2010.07.003
Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses
Abstract
Purpose: Recently, there has been a growing interest in operating medical linear accelerators without a flattening filter. Due to reduced scatter, leaf transmission and radiation head leakage a reduction of out-of-field dose is expected for flattening filter free beams. The aim of the present study was to determine the impact of unflattened beams on peripheral dose for advanced treatment techniques with a large number of MUs.
Material and methods: An Elekta Precise linac was modified to provide 6 and 10 MV photon beams without a flattening filter. Basic beam data were collected and implemented into the TPS Oncentra Masterplan (Nucletron). Leakage radiation, which predominantly contributes to peripheral dose at larger distances from the field edge, was measured using a Farmer type ionisation chamber. SBRT (lung) and IMRT (prostate, head&neck) treatment plans were generated for 6 and 10 MV for both flattened and unflattened beams. All treatment plans were delivered to the relevant anatomic region of an anthropomorphic phantom which was extended by a solid water slab phantom. Dosimetric measurements were performed with TLD-700 rods, radiochromic films and a Farmer type ionisation chamber. The detectors were placed within the slab phantom and positioned along the isocentric longitudinal axis.
Results: Using unflattened beams results in a reduction of treatment head leakage by 52% for 6 and 65% for 10 MV. Thus, peripheral doses were in general smaller for treatment plans calculated with unflattened beams. At about 20 cm distance from the field edge the dose was on average reduced by 23 and 31% for the 6 and 10 MV SBRT plans. For the IMRT plans (10 MV) the average reduction was 16% for the prostate and 18% for the head&neck case, respectively. For all examined cases, the relative deviation between peripheral doses of flattened and unflattened beams was found to increase with increasing distance from the field.
Conclusions: Removing the flattening filter lead to reduced peripheral doses for advanced treatment techniques. The relative difference between peripheral doses of flattened and unflattened beams was more pronounced when the nominal beam energy was increased. Patients may benefit by decreased exposure of normal tissue to scattered dose outside the field.
Copyright © 2010. Published by Elsevier GmbH.
Similar articles
-
Dosimetric characteristics of 6 and 10MV unflattened photon beams.Radiother Oncol. 2009 Oct;93(1):141-6. doi: 10.1016/j.radonc.2009.06.008. Epub 2009 Jul 9. Radiother Oncol. 2009. PMID: 19592123
-
Treatment-planning study of prostate cancer intensity-modulated radiotherapy with a Varian Clinac operated without a flattening filter.Int J Radiat Oncol Biol Phys. 2007 Aug 1;68(5):1567-71. doi: 10.1016/j.ijrobp.2007.04.025. Epub 2007 Jun 4. Int J Radiat Oncol Biol Phys. 2007. PMID: 17544596
-
Treatment planning and delivery of IMRT using 6 and 18MV photon beams without flattening filter.Appl Radiat Isot. 2009 Sep;67(9):1629-37. doi: 10.1016/j.apradiso.2009.03.014. Epub 2009 Mar 25. Appl Radiat Isot. 2009. PMID: 19369083
-
Current status and future perspective of flattening filter free photon beams.Med Phys. 2011 Mar;38(3):1280-93. doi: 10.1118/1.3554643. Med Phys. 2011. PMID: 21520840 Review.
-
[A Proposal for the Absorbed Dose to Water Dosimetry for Flattening Filter-free Beams].Igaku Butsuri. 2016;36(2):79-84. doi: 10.11323/jjmp.36.2_79. Igaku Butsuri. 2016. PMID: 28428458 Review. Japanese.
Cited by
-
Second Cancer Risk after simultaneous integrated boost radiation therapy of right sided breast cancer with and without flattening filter.Strahlenther Onkol. 2016 Oct;192(10):687-95. doi: 10.1007/s00066-016-1025-5. Epub 2016 Aug 17. Strahlenther Onkol. 2016. PMID: 27534409 English.
-
Volumetric modulated arc therapy with flattening filter free beams for isolated abdominal/pelvic lymph nodes: report of dosimetric and early clinical results in oligometastatic patients.Radiat Oncol. 2012 Dec 5;7:204. doi: 10.1186/1748-717X-7-204. Radiat Oncol. 2012. PMID: 23216821 Free PMC article.
-
Impact of flattening-filter-free radiation on the clonogenic survival of astrocytic cell lines.Strahlenther Onkol. 2015 Jul;191(7):590-6. doi: 10.1007/s00066-015-0823-5. Epub 2015 Mar 13. Strahlenther Onkol. 2015. PMID: 25764245
-
Establishing stereotactic body radiotherapy with flattening filter free techniques in the treatment of pulmonary lesions - initial experiences from a single institution.Radiat Oncol. 2016 Jun 8;11:80. doi: 10.1186/s13014-016-0648-0. Radiat Oncol. 2016. PMID: 27411832 Free PMC article. Clinical Trial.
-
Dosimetric differences in flattened and flattening filter-free beam treatment plans.J Med Phys. 2016 Apr-Jun;41(2):92-9. doi: 10.4103/0971-6203.181636. J Med Phys. 2016. PMID: 27217620 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources