Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry
- PMID: 20921437
- DOI: 10.1161/CIRCULATIONAHA.110.976076
Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry
Abstract
Background: Prospective studies defining the risk associated with pacemaker or implantable cardioverter-defibrillator replacement surgeries do not exist. These procedures are generally considered low risk despite results from recent retrospective series reporting higher rates.
Methods and results: We prospectively assessed predefined procedure-related complication rates associated with elective pacemaker or implantable cardioverter-defibrillator generator replacements over 6 months of follow-up. Two groups were studied: those without (cohort 1) and those with (cohort 2) a planned transvenous lead addition for replacement or upgrade to a device capable of additional therapies. Complications were adjudicated by an independent events committee. Seventy-two US academic and private practice centers participated. Major complications occurred in 4.0% (95% confidence interval, 2.9 to 5.4) of 1031 cohort 1 patients and 15.3% (95% confidence interval, 12.7 to 18.1) of 713 cohort 2 patients. In both cohorts, major complications were higher with implantable cardioverter-defibrillator compared with pacemaker generator replacements. Complications were highest in patients who had an upgrade to or a revised cardiac resynchronization therapy device (18.7%; 95% confidence interval, 15.1 to 22.6). No periprocedural deaths occurred in either cohort, although 8 later procedure-related deaths occurred in cohort 2. The 6-month infection rates were 1.4% (95% confidence interval, 0.7 to 2.3) and 1.1% (95% confidence interval, 0.5 to 2.2) for cohorts 1 and 2, respectively.
Conclusions: Pacemaker and implantable cardioverter-defibrillator generator replacements are associated with a notable complication risk, particularly those with lead additions. These data support careful decision making before device replacement, when managing device advisories, and when considering upgrades to more complex systems.
Trial registration: ClinicalTrials.gov NCT00395447.
Similar articles
-
Rate, causes, and impact on patient outcome of implantable device complications requiring surgical revision: large population survey from two centres in Italy.Europace. 2013 Apr;15(4):531-40. doi: 10.1093/europace/eus337. Epub 2013 Feb 13. Europace. 2013. PMID: 23407627
-
Predictors of short-term complications after implantable cardioverter-defibrillator replacement: results from the Ontario ICD Database.Circ Arrhythm Electrophysiol. 2011 Apr;4(2):136-42. doi: 10.1161/CIRCEP.110.959791. Epub 2011 Feb 15. Circ Arrhythm Electrophysiol. 2011. PMID: 21325209
-
Explantation of implantable defibrillator leads using open heart surgery or percutaneous techniques.Ann Thorac Surg. 2008 Jan;85(1):50-5. doi: 10.1016/j.athoracsur.2007.03.048. Ann Thorac Surg. 2008. PMID: 18154777
-
Should recent defibrillator and lead advisories affect decisions to refer patients for implantable cardioverter-defibrillator therapy?Curr Opin Cardiol. 2010 Jan;25(1):23-8. doi: 10.1097/HCO.0b013e328333d375. Curr Opin Cardiol. 2010. PMID: 19907318 Review.
-
Implications of pacemakers and implantable cardioverter defibrillators in urological practice.J Urol. 2011 Oct;186(4):1198-205. doi: 10.1016/j.juro.2011.02.2697. Epub 2011 Aug 16. J Urol. 2011. PMID: 21849189 Review.
Cited by
-
Predictors of Early Cardiac Implantable Electronic Device Lead Dislodgement in the Elderly.Int J Environ Res Public Health. 2022 Nov 10;19(22):14766. doi: 10.3390/ijerph192214766. Int J Environ Res Public Health. 2022. PMID: 36429483 Free PMC article.
-
Variations in surgical technique for permanent pacemaker implantation in West Midlands.Br J Cardiol. 2021 Nov 1;28(1):6. doi: 10.5837/bjc.2021.006. eCollection 2021. Br J Cardiol. 2021. PMID: 35747483 Free PMC article. No abstract available.
-
Leadless intracardiac pacemaker implantation in patients with bradyarrhythmias after spinal cord injury.HeartRhythm Case Rep. 2021 Jul 8;7(10):669-673. doi: 10.1016/j.hrcr.2021.06.013. eCollection 2021 Oct. HeartRhythm Case Rep. 2021. PMID: 34712562 Free PMC article. No abstract available.
-
Longevity decoded: Insights from power consumption analyses into device construction and their clinical implications.Pacing Clin Electrophysiol. 2019 Apr;42(4):407-422. doi: 10.1111/pace.13642. Epub 2019 Mar 5. Pacing Clin Electrophysiol. 2019. PMID: 30802980 Free PMC article.
-
MRI-conditional pacemakers: current perspectives.Med Devices (Auckl). 2014 May 7;7:115-24. doi: 10.2147/MDER.S44063. eCollection 2014. Med Devices (Auckl). 2014. PMID: 24851058 Free PMC article. Review.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Medical