Falsely decreased human chorionic gonadotropin (hCG) results due to increased concentrations of the free beta subunit and the beta core fragment in quantitative hCG assays
- PMID: 20930131
- DOI: 10.1373/clinchem.2010.143479
Falsely decreased human chorionic gonadotropin (hCG) results due to increased concentrations of the free beta subunit and the beta core fragment in quantitative hCG assays
Abstract
Background: Earlier studies have shown that increased concentrations of certain human chorionic gonadotropin (hCG) variants can cause false-negative results in some qualitative hCG devices. The objective of this study was to determine if increased concentrations of hCGβ and hCGβ core fragment (hCGβcf) cause falsely decreased results on 9 commercially available quantitative hCG assays.
Methods: Several concentrations of purified hCGβ and hCGβcf were added to 2 sets of 6 serum samples with and without a fixed concentration of intact hCG. We examined 9 widely used immunoassays to measure immunoreactive hCG. Falsely decreased results were defined as those in which the measured hCG concentration was ≤50% of expected.
Results: High concentrations of hCGβ (≥240 000 pmol/L) produced falsely decreased hCG measurements in 2 assays known to detect this variant. Similarly, high concentrations of hCGβcf (≥63 000 pmol/L) produced falsely decreased hCG measurements in 3 assays that do not detect purified hCGβcf. Two assays were identified that detected both hCGβ and hCGβcf, and neither produced falsely decreased results in the presence of high concentrations of these variants.
Conclusions: Extremely high concentrations of hCG variants can cause falsely decreased results in certain quantitative hCG assays. Of the 9 assays examined, none exhibited falsely decreased results in the presence of hCGβ concentrations typically associated with hCGβ-producing malignancies. Two assays exhibited decreased (>50%) hCG results in the presence of hCGβcf concentrations found during normal pregnancy.
Similar articles
-
Urinary human chorionic gonadotropin isoform concentrations in doping control samples.Drug Test Anal. 2016 Nov;8(11-12):1147-1151. doi: 10.1002/dta.2061. Epub 2016 Sep 29. Drug Test Anal. 2016. PMID: 27594536
-
Using a simulation model to assess risk of false negative point-of-care urinary human chorionic gonadotropin device results due to high-dose hook interference.Clin Biochem. 2015 Feb;48(3):99-104. doi: 10.1016/j.clinbiochem.2014.11.004. Epub 2014 Nov 14. Clin Biochem. 2015. PMID: 25448030
-
Estimating the hCGβcf in urine during pregnancy.Clin Biochem. 2016 Feb;49(3):282-6. doi: 10.1016/j.clinbiochem.2015.10.020. Epub 2015 Nov 2. Clin Biochem. 2016. PMID: 26541857
-
Determination of human chorionic gonadotropin.Best Pract Res Clin Endocrinol Metab. 2013 Dec;27(6):783-93. doi: 10.1016/j.beem.2013.10.005. Epub 2013 Oct 26. Best Pract Res Clin Endocrinol Metab. 2013. PMID: 24275190 Review.
-
Human chorionic gonadotropin in pregnancy diagnostics.Clin Chim Acta. 2011 Aug 17;412(17-18):1515-20. doi: 10.1016/j.cca.2011.05.025. Epub 2011 May 25. Clin Chim Acta. 2011. PMID: 21635878 Review.
Cited by
-
Abnormally low hCG in a complete hydatidiform molar pregnancy: The hook effect.Gynecol Oncol Case Rep. 2011 Oct 20;1(1):6-7. doi: 10.1016/j.gynor.2011.10.003. eCollection 2011. Gynecol Oncol Case Rep. 2011. PMID: 24371588 Free PMC article.
-
Immunoextraction-tandem mass spectrometry method for measuring intact human chorionic gonadotropin, free β-subunit, and β-subunit core fragment in urine.Clin Chem. 2014 Aug;60(8):1089-97. doi: 10.1373/clinchem.2014.222703. Epub 2014 Jun 4. Clin Chem. 2014. PMID: 24899693 Free PMC article.
-
Candidate epitopes for measurement of hCG and related molecules: the second ISOBM TD-7 workshop.Tumour Biol. 2013 Dec;34(6):4033-57. doi: 10.1007/s13277-013-0994-6. Epub 2013 Sep 26. Tumour Biol. 2013. PMID: 24068570 Free PMC article.
-
False-negative qualitative human chorionic gonadotropin (hCG) test result ('hook effect') with classical ultrasound findings of complete molar pregnancy: an uncommon case.Oxf Med Case Reports. 2024 Jan 27;2024(1):omad147. doi: 10.1093/omcr/omad147. eCollection 2024 Jan. Oxf Med Case Reports. 2024. PMID: 38292164 Free PMC article.
-
Analytical and clinical validation of the Immulite 1000 hCG assay for quantitative analysis in urine.Clin Chim Acta. 2013 Jun 5;421:104-8. doi: 10.1016/j.cca.2013.02.026. Epub 2013 Mar 5. Clin Chim Acta. 2013. PMID: 23470427 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources