Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening

Affiliations
Free PMC article
Comparative Study

Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening

Boel Bengtsson et al. Acta Ophthalmol. 2012 Jun.
Free PMC article

Abstract

Purpose: To investigate the measures of validity for selective or population screening of the time-domain Stratus and the spectral-domain Cirrus Optical Coherence Tomography (OCT) imaging techniques in a population-based sample and in clinical glaucoma patients at different stages of glaucoma.

Patients and methods: A random sample of 307 subjects living in two rural areas in southern Sweden, and a random sample of 394 clinical glaucoma patients were selected. A large battery of examinations, including Stratus and Cirrus OCT was performed. OCT retinal nerve fibre layer (RNFL) thickness analyses for average thickness, quadrant and clock-hour sectors were compared with normative significance limits available in the instruments.

Results: The population-based sample included 129 healthy and nine glaucoma subjects, and the sample of clinical glaucoma patients included 138 patients. Specificity and positive predictive values were generally better with Stratus than for Cirrus, and sensitivity was generally better with Cirrus. With the average RNFL thickness parameter, Stratus reached 100% specificity and a positive predictive value of 100% and 68% sensitivity in the whole group of the clinical glaucoma patients, but sensitivity was only 28% among the earliest stage glaucoma patients. Sensitivity increased considerably when relying on the quadrant sector parameter, while specificity decreased only marginally.

Conclusion: Stratus, with high specificity and positive predictive values, seemed to be best of choice for screening purposes, while Cirrus, with high sensitivity, was the better choice for early detection.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Mean Deviation (MD) values of visual fields assessed in clinical glaucoma patients (dark bars) and in the nine subjects in the population-based sample who were found to be glaucomatous (white bars). Seven of the latter nine patients had mild visual field defects with MD values better than −6 dB, and two subjects, both previously diagnosed, had MD values of approximately −15 dB.

References

    1. Bengtsson B. The prevalence of glaucoma. Br J Ophthalmol. 1981;65:46–49. - PMC - PubMed
    1. Budenz DL, Michael A, Chang RT, McSoley J, Katz J. Sensitivity and specificity of the Stratus OCT for perimetric glaucoma. Ophthalmology. 2005;112:3–9. - PubMed
    1. Budenz DL, Anderson DR, Varma R, et al. Determinants of normal retinal nerve fiber layer thickness measured by Stratus OCT. Ophthalmology. 2007;114:1046–1052. - PMC - PubMed
    1. Chang RT, Knight OJ, Feuer WJ, Budenz DL. Sensitivity and specificity of time-domain versus spectral-domain optical coherence tomography in diagnosing early to moderate glaucoma. Ophthalmology. 2009;116:2294–2299. - PubMed
    1. Chen H-Y, Huang M-L. Discrimination between normal and glaucomatous eyes using Stratus optical coherence tomography in Taiwan Chinese subjects. Graefe's Arch Clin Exp Ophthalmol. 2005;243:894–902. - PubMed

MeSH terms