False negative and positive motor evoked potentials in one patient: is single motor evoked potential monitoring reliable method? A case report and literature review
- PMID: 20956881
- DOI: 10.1097/BRS.0b013e3181d8fabb
False negative and positive motor evoked potentials in one patient: is single motor evoked potential monitoring reliable method? A case report and literature review
Abstract
Study design: A case report and literature review.
Objective: To report a false negative and delayed positive motor-evoked potential (MEP) in 1 patient.
Summary of background data: An unreliable MEP can result in fatal outcomes because surgeons have recently begun to depend on the MEP for intraoperative decision-making.
Methods: We report a case of a false MEP during scoliosis surgery that showed false negative and positive MEPs during a series of operations.
Results: A 23-year-old man with a history of spondyloepiphyseal dysplasia presented with severe kyphoscoliosis. The initial neurologic examination did not reveal any neurologic abnormalities. Surgical correction and fusion were performed with transcranial MEP monitoring. During the entire procedure, the MEP did not reveal any signs of cord injury. However, lower limb paralysis and paresthesia was observed when the patient awakened. After 2 additional surgical procedures to recover the neurologic deficit, the MEP did not show any signs of cord injury but the patient's neurologic status had recovered slightly. At postoperative day 8, the neurologic status recovered, and a third operation was performed to fix the long rods. However, there were abnormal amplitudes in both lower limbs but the patient's neurologic status was almost normal.
Conclusion: From our experience of false negative and positive MEP in 1 patient, it is concluded that undesirable events can occur with use of MEP in scoliosis or other spinal surgery. Therefore, we warn the surgeons too heavily rely on the MEP monitoring, and propose a further prospective study as well as use of alternative method that can improve the reliability of single MEP.
Similar articles
-
Evaluation of reliability of post-tetanic motor-evoked potential monitoring during spinal surgery under general anesthesia.Spine (Phila Pa 1976). 2008 Dec 15;33(26):E994-E1000. doi: 10.1097/BRS.0b013e318188adfc. Spine (Phila Pa 1976). 2008. PMID: 19092611
-
The efficacy of motor evoked potentials in fixed sagittal imbalance deformity correction surgery.Spine (Phila Pa 1976). 2008 Jun 1;33(13):E414-24. doi: 10.1097/BRS.0b013e318175c292. Spine (Phila Pa 1976). 2008. PMID: 18520928
-
Transcranial motor-evoked potentials combined with response recording through compound muscle action potential as the sole modality of spinal cord monitoring in spinal deformity surgery.Spine (Phila Pa 1976). 2008 May 1;33(10):1100-6. doi: 10.1097/BRS.0b013e31816f5f09. Spine (Phila Pa 1976). 2008. PMID: 18449044
-
False-negative transcranial motor-evoked potentials during scoliosis surgery causing paralysis: a case report with literature review.Spine (Phila Pa 1976). 2009 Nov 15;34(24):E896-900. doi: 10.1097/BRS.0b013e3181b40d4f. Spine (Phila Pa 1976). 2009. PMID: 19910760 Review.
-
[Intraoperative neurophysiological monitoring of the spinal cord].Masui. 2012 Jan;61(1):16-24. Masui. 2012. PMID: 22338856 Review. Japanese.
Cited by
-
Transient neurological deficit following midthoracic decompression for severe stenosis: a series of three cases.Eur Spine J. 2013 Sep;22(9):2057-61. doi: 10.1007/s00586-013-2829-y. Epub 2013 May 14. Eur Spine J. 2013. PMID: 23670822 Free PMC article.
-
Intraoperative electrophysiological monitoring during posterior craniocervical distraction and realignment for congenital craniocervical anomaly.Eur Spine J. 2015 Apr;24(4):671-8. doi: 10.1007/s00586-015-3791-7. Epub 2015 Feb 7. Eur Spine J. 2015. PMID: 25663645
-
Optical monitoring and detection of spinal cord ischemia.PLoS One. 2013 Dec 16;8(12):e83370. doi: 10.1371/journal.pone.0083370. eCollection 2013. PLoS One. 2013. PMID: 24358279 Free PMC article.
-
Laser safety in fiber-optic monitoring of spinal cord hemodynamics: a preclinical evaluation.J Biomed Opt. 2018 Jun;23(6):1-9. doi: 10.1117/1.JBO.23.6.065003. J Biomed Opt. 2018. PMID: 29923371 Free PMC article.
-
Change in body surface temperature as an ancillary measurement to motor evoked potentials.Spinal Cord. 2015 Nov;53(11):827-34. doi: 10.1038/sc.2015.90. Epub 2015 Jun 30. Spinal Cord. 2015. PMID: 26123207
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials