Orthodontic Bracket Manufacturing Tolerances and Dimensional Differences between Select Self-Ligating Brackets
- PMID: 20981299
- PMCID: PMC2958443
- DOI: 10.4061/2010/781321
Orthodontic Bracket Manufacturing Tolerances and Dimensional Differences between Select Self-Ligating Brackets
Abstract
In all manufacturing processes there are tolerances; however, orthodontic bracket manufacturers seldom state the slot dimensional tolerances. This experiment develops a novel method of analyzing slot profile dimensions using photographs of the slot. Five points are selected along each wall, and lines are fitted to define a trapezoidal slot shape. This investigation measures slot height at the slot's top and bottom, angles between walls, slot taper, and the linearity of each wall. Slot dimensions for 30 upper right central incisor self-ligating stainless steel brackets from three manufacturers were evaluated. Speed brackets have a slot height 2% smaller than the nominal 0.559 mm size and have a slightly convergent taper. In-Ovation brackets have a divergent taper at an average angle of 1.47 degrees. In-Ovation is closest to the nominal value of slot height at the slot base and has the smallest manufacturing tolerances. Damon Q brackets are the most rectangular in shape, with nearly 90-degree corners between the slot bottom and walls. Damon slot height is on average 3% oversized.
Figures





References
-
- Kalpakjian S, Schmid SR. Manufacturing Engineering and Technology. 4th edition. Upper Saddle River, NJ, USA: Prentice Hall; 2001.
-
- Fischer-Brandies H, Orthuber W, Es-Souni M, Meyer S. Torque transmission between square wire and bracket as a function of measurement, form and hardness parameters. Journal of Orofacial Orthopedics. 2000;61(4):258–265. - PubMed
-
- Gioka C, Eliades T. Materials-induced variation in the torque expression of preadjusted appliances. American Journal of Orthodontics and Dentofacial Orthopedics. 2004;125(3):323–328. - PubMed
-
- Meling TR, Ødegaard J, Seqner D. On bracket slot height: a methodologic study. American Journal of Orthodontics and Dentofacial Orthopedics. 1998;113(4):387–393. - PubMed
-
- Meling T, Ødegaard J, Holthe K, Meling EO, Segner D. A formula for the displacement of an arch wire when subjected to a second-order couple. American Journal of Orthodontics and Dentofacial Orthopedics. 1998;113(6):632–640. - PubMed