Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants
- PMID: 2099293
Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants
Abstract
A new electronic root canal measuring device called "ENDEX" which is able to determine the root canal length under moistened conditions using a relative value of different alternative electric currents is discussed in this study. This study is designed to investigate the efficiencies of the new device, in vitro, with respect to change in the root canal environment; namely the influence of the size of K-file, the diameter of the apical foramen and the various root canal irrigants. According to the present study, it was confirmed that the location of the apical foramen was clearly detected under moistened root canal with saline solution using any size of K-file. The locations of the tip of K-files were 0.5 mm inside the apical foramen when the relative value displayed 0.3 Volt, so that the apical constriction could not disturb during root canal measurement. One of the big problem with prior electrical measurement devices was not being able to determine the canal length under moistened conditions. However with this improved device, intracanal irrigants, such as saline, 5% NaOC1, 14% EDTA and 3% H2O2 did not interfere in detecting apical foramen irregardless of the size of K-file and the size of the apical foramen.
Similar articles
-
Evaluation of accuracy of an electric device (Neosono D-SE) for the measurement of tooth length.Indian J Dent Res. 1989 Apr-Sep;1(2-3):58-65. Indian J Dent Res. 1989. PMID: 2490122
-
[Accuracy of 2 electronic canal length measuring instruments].Rev Fr Endod. 1991 Mar;10(1):27-33. Rev Fr Endod. 1991. PMID: 2020777 Review. French.
-
New electronic canal measuring device based on the ratio method.J Endod. 1994 Mar;20(3):111-4. doi: 10.1016/S0099-2399(06)80053-1. J Endod. 1994. PMID: 7996080
-
[The evaluation of the ability of Apex D.S.P. to determine the length of root canal. In vitro study].Ann Acad Med Stetin. 2008;54(2):37-40. Ann Acad Med Stetin. 2008. PMID: 19374229 Polish.
-
The fundamental operating principles of electronic root canal length measurement devices.Int Endod J. 2006 Aug;39(8):595-609. doi: 10.1111/j.1365-2591.2006.01131.x. Int Endod J. 2006. PMID: 16872454 Review.
Cited by
-
Accuracy and stability of electronic apex locator length measurements in root canals with wide apical foramen: an ex vivo study.BDJ Open. 2020 Nov 17;6(1):22. doi: 10.1038/s41405-020-00052-3. BDJ Open. 2020. PMID: 33298851 Free PMC article.
-
Working Length Determination Using Cone-Beam Computed Tomography, Periapical Radiography and Electronic Apex Locator in Teeth with Apical Periodontitis: A Clinical Study.Iran Endod J. 2016 Summer;11(3):164-8. doi: 10.7508/iej.2016.03.003. Epub 2016 May 1. Iran Endod J. 2016. PMID: 27471524 Free PMC article.
-
Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator.Restor Dent Endod. 2015 Feb;40(1):44-9. doi: 10.5395/rde.2015.40.1.44. Epub 2014 Nov 5. Restor Dent Endod. 2015. PMID: 25671212 Free PMC article.
-
Accuracy of two different apex locators in primary teeth with and without root resorption.Clin Oral Investig. 2008 Jun;12(2):137-41. doi: 10.1007/s00784-007-0157-5. Epub 2007 Nov 20. Clin Oral Investig. 2008. PMID: 18027005
-
Use of artificial primary teeth for endodontic laboratory research: experiments related to canal length determination.BMC Oral Health. 2017 Nov 17;17(1):131. doi: 10.1186/s12903-017-0420-3. BMC Oral Health. 2017. PMID: 29149844 Free PMC article.