In vitro performance comparison of the Sensormedics 3100A and B high-frequency oscillatory ventilators
- PMID: 21037502
- DOI: 10.1097/PCC.0b013e3181fe3028
In vitro performance comparison of the Sensormedics 3100A and B high-frequency oscillatory ventilators
Abstract
Objective: The Sensormedics 3100A and 3100B are widely used to provide high-frequency oscillatory ventilation in clinical practice. Infants and children <35 kg are typically oscillated with the 3100A and >35 kg with the 3100B. This study compares the effect of ventilator and patient parameters on delivered tidal volume during high-frequency oscillatory ventilation of a test lung with these devices.
Design: Laboratory-based study.
Subjects: Test lung and Sensormedics 3100A and 3100B high-frequency oscillators.
Interventions: A previously validated hot-wire flowmeter (Florian) was placed in series with either a 3100A (n = 3) or 3100B (n = 3) ventilator and a Michigan test lung. Tidal volumes were measured over a range of mean airway pressure, inspiratory:expiratory ratio, frequency, pressure amplitude, and endotracheal tube internal diameter.
Measurements and main results: The 3100A and 3100B delivered similar tidal volumes across a range of ventilator parameters for an inspiratory:expiratory ratio of 1:1, differing by <10%. However, at an inspiratory:expiratory ratio of 1:2, there was a statistically significant decrease in tidal volume for the 3100B compared with the 3100A at lower frequencies, which was partially mitigated by increasing pressure amplitude. The difference in the generated pressure and flow waveforms may account for the observed tidal volume differences between the high-frequency oscillatory ventilation models. Delivered tidal volume was highly dependent on endotracheal tube size.
Conclusions: Multiple variables contribute to the delivered tidal volume during high-frequency oscillatory ventilation, including ventilator model selection and endotracheal tube size. It is possible that real-time, clinical monitoring of delivered tidal volume during high-frequency oscillatory ventilation would allow better titration and maximize performance of these ventilators in caring for critically ill patients.
Similar articles
-
Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome.Crit Care Med. 2007 Jun;35(6):1522-9. doi: 10.1097/01.CCM.0000266586.04676.55. Crit Care Med. 2007. PMID: 17440422
-
Four methods of measuring tidal volume during high-frequency oscillatory ventilation.Crit Care Med. 2006 Mar;34(3):751-7. doi: 10.1097/01.CCM.0000201400.63304.41. Crit Care Med. 2006. PMID: 16505661
-
Reliable tidal volume estimates at the airway opening with an infant monitor during high-frequency oscillatory ventilation.Crit Care Med. 2001 Oct;29(10):1925-30. doi: 10.1097/00003246-200110000-00013. Crit Care Med. 2001. PMID: 11588453
-
A protocol for high-frequency oscillatory ventilation in adults: results from a roundtable discussion.Crit Care Med. 2007 Jul;35(7):1649-54. doi: 10.1097/01.CCM.0000269026.40739.2E. Crit Care Med. 2007. PMID: 17522576
-
Clinical application of ventilator modes: Ventilatory strategies for lung protection.Aust Crit Care. 2010 May;23(2):71-80. doi: 10.1016/j.aucc.2010.03.003. Epub 2010 Apr 7. Aust Crit Care. 2010. PMID: 20378369 Review.
Cited by
-
Searching the right way to treat neonatal abstinence syndrome*.Pediatr Crit Care Med. 2014 Feb;15(2):175-6. doi: 10.1097/PCC.0000000000000037. Pediatr Crit Care Med. 2014. PMID: 24492187 Free PMC article.
-
The Physiological Basis of High-Frequency Oscillatory Ventilation and Current Evidence in Adults and Children: A Narrative Review.Front Physiol. 2022 Apr 26;13:813478. doi: 10.3389/fphys.2022.813478. eCollection 2022. Front Physiol. 2022. PMID: 35557962 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources