Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics
- PMID: 21039345
- DOI: 10.1111/j.1399-6576.2010.02321.x
Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics
Abstract
Background: A commercially available pulse oximeter that reports blood haemoglobin (Hb) concentration is evaluated. This study considers whether this device can provide serial Hb data that would be sufficiently reliable for volume kinetic analysis of infusion fluids.
Methods: Forty infusions of 5 or 10 ml/kg of acetated Ringer's solution were given over 15 min in 10 healthy volunteers. Hb was measured on 17 different occasions over 120 min using the Radical 7 pulse oximeter and compared with the result of invasive blood sampling (control). A one-volume kinetic model was applied to each data series. The pulse oximeter also reported the perfusion index (PI).
Results: The median deviation between the 680 invasive and non-invasive Hb samples (the accuracy) was 1.6% and the absolute median deviation (precision) was 4.6%. Between-subject factors explained half of the variation in the difference between non-invasive vs. invasive sampling. Ten of the 40 non-invasive series of Hb values were discarded from kinetic analysis due to poor quality. The remaining 30 series showed a smaller distribution volume for the infused fluid when kinetic analysis was based on the non-invasive method (3.0 vs. 5.3l; P<0.001). This was due to co-variance with the PI, which exaggerated the decrease in Hb caused by the infusions. The non-invasive method might provide useful kinetic data at the group level, but individual curves deviated too much from the invasive data to be reliable.
Conclusions: Non-invasive measurement of the Hb concentration during volume loading could not provide useful kinetic data for individuals.
© 2010 The Authors. Journal compilation © 2010 The Acta Anaesthesiologica Scandinavica Foundation.
Comment in
-
Instability in non-invasive haemoglobin measurement: a possible influence of oxygen administration.Acta Anaesthesiol Scand. 2011 Aug;55(7):902. doi: 10.1111/j.1399-6576.2011.02469.x. Epub 2011 Jun 9. Acta Anaesthesiol Scand. 2011. PMID: 21658016 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources