Repeated transchoroidal implantation and explantation of compound subretinal prostheses: an exploratory study in rabbits
- PMID: 21052911
- DOI: 10.1007/s10384-010-0851-8
Repeated transchoroidal implantation and explantation of compound subretinal prostheses: an exploratory study in rabbits
Abstract
Purpose: For human trials with retinal prostheses it is mandatory to develop procedures to safely explant and possibly reimplant the devices. This prompted us to investigate in a small exploratory study the safety of repeated transchoroidal implantation and explantation procedures of complex subretinal devices in laboratory animals.
Methods: Repeated transchoroidal surgery was performed in four rabbits. The rabbits were examined by clinical examination and funduscopy. Function was assessed by electroretinography and cortical recordings following light and subretinal electrical stimulation. Sections of the retina and of the implantation channel were examined by light microscopy.
Results: Using the same access route, repeated transchoroidal subretinal implantation surgery was successfully performed in all cases. Fixation of implants was stable for up to 13 months; retinas remained attached at all examination dates. Electroretinograms and visual evoked cortical potential proved retinal and visual pathway integrity. Subretinal electrical stimulation elicited retinal and cortical responses. While retinal morphology at earlier stages was found to be essentially unaltered, atrophic disorganization in the region of the subretinal channel was observed after 10 months and after subretinal electrical stimulation.
Conclusions: Repeated transchoroidal surgery can be safely performed for implantation, explantation, and reimplantation of subretinal devices in rabbits. With modifications, we believe the technique can be applied in human surgery.
Similar articles
-
Subretinal electrical stimulation of the rabbit retina with acutely implanted electrode arrays.Graefes Arch Clin Exp Ophthalmol. 2004 Jul;242(7):587-96. doi: 10.1007/s00417-004-0862-6. Epub 2004 Jun 5. Graefes Arch Clin Exp Ophthalmol. 2004. PMID: 15197555
-
Studies on the feasibility of a subretinal visual prosthesis: data from Yucatan micropig and rabbit.Graefes Arch Clin Exp Ophthalmol. 2001 Dec;239(12):961-7. doi: 10.1007/s004170100368. Graefes Arch Clin Exp Ophthalmol. 2001. PMID: 11820703
-
Subretinal implantation and testing of polyimide film electrodes in cats.Graefes Arch Clin Exp Ophthalmol. 2005 May;243(5):464-8. doi: 10.1007/s00417-004-1049-x. Epub 2004 Dec 1. Graefes Arch Clin Exp Ophthalmol. 2005. PMID: 15578200
-
[Status of the subretinal implant project. An overview].Ophthalmologe. 2005 Oct;102(10):941-9. doi: 10.1007/s00347-005-1258-7. Ophthalmologe. 2005. PMID: 16151772 Review. German.
-
[Research progress of subretinal implant based on electronic stimulation].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008 Feb;25(1):212-4, 219. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008. PMID: 18435293 Review. Chinese.
Cited by
-
Acute Rabbit Eye Model for Testing Subretinal Prostheses.Transl Vis Sci Technol. 2019 Oct 2;8(5):20. doi: 10.1167/tvst.8.5.20. eCollection 2019 Sep. Transl Vis Sci Technol. 2019. PMID: 31602345 Free PMC article.
-
Angiographic findings following tack fixation of a wireless epiretinal retina implant device in blind RP patients.Graefes Arch Clin Exp Ophthalmol. 2011 Sep;249(9):1281-6. doi: 10.1007/s00417-011-1653-5. Epub 2011 Apr 5. Graefes Arch Clin Exp Ophthalmol. 2011. PMID: 21465287 Clinical Trial.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources