. 2010 Nov-Dec;45(6):642-4.
doi: 10.4085/1062-6050-45.6.642.

Temperature measurement reliability and validity with thermocouple extension leads or changing lead temperature

Affiliations

Temperature measurement reliability and validity with thermocouple extension leads or changing lead temperature

Lisa S Jutte et al. J Athl Train. 2010 Nov-Dec.

Abstract

Context: Thermocouples' leads are often too short, necessitating the use of an extension lead.

Objective: To determine if temperature measures were influenced by extension-lead use or lead temperature changes.

Design: Descriptive laboratory study.

Setting: Laboratory.

Other participants: Experiment 1: 10 IT-21 thermocouples and 5 extension leads. Experiment 2: 5 IT-21 and PT-6 thermocouples.

Methods: In experiment 1, temperature data were collected on 10 IT-21 thermocouples in a stable water bath with and without extension leads. In experiment 2, temperature data were collected on 5 IT-21 and PT-6 thermocouples in a stable water bath before, during, and after ice-pack application to extension leads.

Results: In experiment 1, extension leads did not influence IT-21 validity (P = .45) or reliability (P = .10). In experiment 2, postapplication IT-21 temperatures were greater than preapplication and application measures (P < .05).

Conclusions: Extension leads had no influence on temperature measures. Ice application to leads may increase measurement error.

PubMed Disclaimer

Figures

Figure
Figure
The distal ends of 10 IT-21 thermocouples were inserted through a polystyrene cooler, secured with a silicone-based polymer, and connected to the Iso-Thermex.

References

    1. Jutte L. S., Knight K. L., Long B. C., Hawkins J. R., Schulthies S. S., Dalley E. B. The uncertainty (validity and reliability) of three electrothermometers in therapeutic modality research. J Athl Train. 2005;40(3):207–210. - PMC - PubMed
    1. Long B. C., Jutte L. S., Knight K. L. Thermocouples interfaced to electrothermometers respond differently when immersed in 5 water-bath temperatures. J Athl Train. 2010;45(4):338–343. - PMC - PubMed
    1. National Institute of Standards and Technology. The NIST reference on constants, units, and uncertainty: glossary. National Institute of Standards and Technology (NIST) Web site. http://physics.nist.gov/cuu/Uncertainty/glossary.html. Updated October 2000. Accessed February 21, 2005.
    1. Physitemp Instruments. Physitemp Precision Temperature Specialists. Clifton, NJ: Physitemp Instruments Inc; 1999. Temperature microprobes.
    1. Jutte L. S., Knight K. L., Long B. C. Reliability and validity of electrothermometers and associated thermocouples. J Sport Rehabil. 2008;17(1):50–59. - PubMed