HIV-1 load comparison using four commercial real-time assays
- PMID: 21068276
- PMCID: PMC3020484
- DOI: 10.1128/JCM.01688-10
HIV-1 load comparison using four commercial real-time assays
Abstract
The HIV-1 RNA viral load is commonly used for the monitoring of disease progression and antiretroviral treatment of HIV-1-infected patients. Since the misestimating of values could lead to inappropriate therapeutical management, the comparative performances, especially the ability to span the genetic diversity of HIV-1, of available automated real-time assays need to be evaluated. We conducted a prospective study with 74 consenting patients enrolled between March 2007 and November 2008. A blood sample was obtained at the time of diagnosis of HIV seropositivity and blindly tested for HIV-1 RNA by at least 4 commercial tests: the Abbott m2000 RealTime HIV-1, bioMérieux NucliSens EasyQ HIV-1, version 1.2 (v1.2), and Cobas AmpliPrep/Cobas TaqMan (CAP/CTM) v1.0 and v2.0 assays. The means of difference were null between CAP/CTM v2.0 and Abbott for CRF02_AG subtypes but positive in favor of CAP/CTM v2.0 for genotype B and negative in favor of NucliSens for all genotypes. The standard deviation (SD) of difference ranged from 0.3 to 0.59, depending on the considered couples of assays. Reliabilities of these four tests, appreciated by the standard deviation of difference between the measurement and the estimated "true" viral load and by the coefficient of reliability, were significantly different (P < 10(-4)) among each other. Significant differences were also observed within each group of HIV-1 genotype. The global disparity was higher for CRF02_AG than for B subtypes. This study indicates a risk of viral load misestimating or discrepancies between techniques, depending on the HIV-1 subtype, and speaks in favor of using the same assay for the monitoring of HIV-1-infected patients.
Figures
References
-
- Altman, D. G., and J. M. Bland. 1983. Measurement in medicine: the analysis of method comparison studies. Statistician 32:307-317.
-
- Damond, F., B. Roquebert, A. Benard, G. Collin, M. Miceli, P. Yeni, F. Brun-Vezinet, and D. Descamps. 2007. Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays. J. Clin. Microbiol. 45:3436-3438. - PMC - PubMed
-
- De Bel, A., D. Marissens, L. Debaisieux, C. Liesnard, S. Van den Winjngaert, S. Lauwers, and D. Piérard. 2010. Correction of underquantification of human immunodeficiency virus type 1 load with the second version of the Roche Cobas AmpliPrep/Cobas TaqMan Assay. J. Clin. Microbiol. 48:1337-1342. - PMC - PubMed
-
- de Mendoza, C., M. Koppelman, B. Montes, V. Ferre, V. Soriano, H. Cuypers, M. Segondy, and T. Oosterlaken. 2005. Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma. J. Virol. Methods 127:54-57. - PubMed