Comparative Study
. 2011 Feb;51(2):126-9.
doi: 10.1007/s00117-010-2078-6.

[Comparison of dignity determination of mammographic microcalcification with two systems for digital full-field mammography with different detector resolution: a retrospective clinical study]

[Article in German]
Affiliations
Comparative Study

[Comparison of dignity determination of mammographic microcalcification with two systems for digital full-field mammography with different detector resolution: a retrospective clinical study]

[Article in German]
R Schulz-Wendtland et al. Radiologe. 2011 Feb.

Abstract

Purpose: The aim of this retrospective clinical study was to compare the diagnostic accuracy of the novel 50 µm FFDM (full-field digital mammography) system (DR) with an established 70 µm system (DR) in the differential diagnosis between benign and malignant clusters of microcalcification (n=50) (BI-RADS™ classification 4/5) and to assess the possible incremental value of the 50 µm pixel-pitch on specificity.

Material and methods: From March 2009 to September 2009, 50 patients underwent full-field digital mammography (FFDM) (detector resolution 70 µm) (Novation, Siemens, Erlangen, Germany). As there were suspicious signs of microcalcification classified with BI-RADS™ 4/5 after diagnosis and preoperative wire localization, control images were made with the new FFDM system (detector: resolution 50 µm) (Amulet, Fujifilm, Tokyo, Japan) with the same exposure parameters. The diagnosis was determined after the operation by five radiologists with different experience in digital mammography from randomly distributed mediolateral views (monitor reading) whose results were correlated with the final histology of all lesions.

Results: Histopathology revealed 19 benign and 31 malignant lesions in 50 patients after open biopsy. The results of the five readers showed a higher sensitivity of the new FFDM system (80.0%) in the ability to recognize malignant microcalcification in comparison to the established system (74.8%). The specificity (75.8 versus 71.6%) was slightly higher for the new system but these results were not statistically significant (p<0.001). Considering the diagnostic accuracy, the new system (detector: resolution 50 µm) was also slightly superior to the well-known system (detector: resolution 70 µm) (80.1% versus 76.4%).

Conclusions: Our study has shown that the new full-field digital mammography system using the novel detector compared with the already established FFDM system with respect to the assessment of microcalcification is at least equivalent.

PubMed Disclaimer

Similar articles

Cited by

References

    1. N Engl J Med. 2005 Oct 27;353(17):1773-83 - PubMed
    1. AJR Am J Roentgenol. 1986 Apr;146(4):661-3 - PubMed
    1. Med Phys. 2009 Nov;36(11):5139-48 - PubMed
    1. Eur J Radiol. 2009 Nov;72(2):258-65 - PubMed
    1. Acta Radiol. 2009 Jan;50(1):3-14 - PubMed

LinkOut - more resources