[First experiments for the detection of simulated mammographic lesions: digital full field mammography with a new detector with a double plate of pure selenium]
- PMID: 21069512
- DOI: 10.1007/s00117-010-2077-7
[First experiments for the detection of simulated mammographic lesions: digital full field mammography with a new detector with a double plate of pure selenium]
Abstract
Purpose: The article describes an experimental phantom study of a system for digital full field mammography with a new digital detector with a double plate of pure selenium.
Materials and methods: The experiments were carried out with the new full field digital mammography system Amulet from FujiFilm. This system has a new detector (18×24 cm(2)) on the basis of highly purified amorphous selenium (a-Se) with a pixel size of 50 µm. The x-rays are converted into electric signals in the first plate which are read into the second plate with the help of an optical switch and demonstrated in the form of an image. In this way a better pixel size/volume and signal-to-noise ratio should be achieved. The object of the investigation was the Wisconsin Mammographic Random Phantom, Model 152 A (Technical Performance Mo/Mo, 28 kV, 100 mAs). Five investigators with different experiences in mammography each received three images on a monitor with different random positions of the simulated lesions in the phantom for assessment. The detection rates were compared under the same conditions with the results of two other full field digital mammography systems.
Results: The median detection rate for all images and investigators for the new doubled plated a-Se detector with optical switch was 98.7%. For both other systems with a-Si or and a-Se detectors the detection rate was 89.8% or 97.3%, respectively. There were no significant differences in the detection rate of the simulated breast lesions for all three systems considering the interobserver and intraobserver variation.
Conclusion: The first phantom study for the detection of simulated breast lesions with the new full field digital mammography system Amulet demonstrates equivalent results with the other systems used in the clinical routine. The trend towards superiority of the new system has to be confirmed in further clinical studies.
Similar articles
-
[Comparison of dignity determination of mammographic microcalcification with two systems for digital full-field mammography with different detector resolution: a retrospective clinical study].Radiologe. 2011 Feb;51(2):126-9. doi: 10.1007/s00117-010-2078-6. Radiologe. 2011. PMID: 21069511 German.
-
Experimental investigations of image quality in X-ray mammography with a conventional screen film system (SFS) and a new full-field digital mammography unit (DR) with a-Se-detector.Rofo. 2003 Jun;175(6):766-8. doi: 10.1055/s-2003-39936. Rofo. 2003. PMID: 12811687
-
Imaging performance of an amorphous selenium digital mammography detector in a breast tomosynthesis system.Med Phys. 2008 May;35(5):1978-87. doi: 10.1118/1.2903425. Med Phys. 2008. PMID: 18561674 Free PMC article.
-
High resolution dual detector volume-of-interest cone beam breast CT--Demonstration with a bench top system.Med Phys. 2011 Dec;38(12):6429-42. doi: 10.1118/1.3656040. Med Phys. 2011. PMID: 22149826 Free PMC article.
-
A comparison between objective and subjective image quality measurements for a full field digital mammography system.Phys Med Biol. 2006 May 21;51(10):2441-63. doi: 10.1088/0031-9155/51/10/006. Epub 2006 Apr 26. Phys Med Biol. 2006. PMID: 16675862
Cited by
-
Pilot Study on the Detection of Simulated Lesions Using a 2D and 3D Digital Full-Field Mammography System with a Newly Developed High Resolution Detector Based on Two Shifts of a-Se.Geburtshilfe Frauenheilkd. 2012 May;72(5):408-411. doi: 10.1055/s-0031-1298158. Geburtshilfe Frauenheilkd. 2012. PMID: 25298545 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical