Bicinchoninic acid (BCA) assay in low volume
- PMID: 21078286
- DOI: 10.1016/j.ab.2010.11.015
Bicinchoninic acid (BCA) assay in low volume
Abstract
The BCA assay is a colorimetric method for estimating protein concentration. In 96-well plates, the relationship between protein content and absorbance is nearly linear over a wide range; however, performance is reduced in lower volume. To overcome this limitation, we performed the BCA assays in opaque, white 384-well plates. These plates emit fluorescence between 450-600 nm when excited at 430 nm; thus, their fluorescence is quenched by the BCA chromophore (λ(max) 562 nm). This arrangement allowed accurate determination of protein content using only 2 μL of sample. Moreover, soluble flourescein could replace the white plates, creating a homogenous format.
Copyright © 2010 Elsevier Inc. All rights reserved.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources