Analog signal multiplexing for PSAPD-based PET detectors: simulation and experimental validation
- PMID: 21081831
- PMCID: PMC3691840
- DOI: 10.1088/0031-9155/55/23/001
Analog signal multiplexing for PSAPD-based PET detectors: simulation and experimental validation
Abstract
A 1 mm(3) resolution clinical positron emission tomography (PET) system employing 4608 position-sensitive avalanche photodiodes (PSAPDs) is under development. This paper describes a detector multiplexing technique that simplifies the readout electronics and reduces the density of the circuit board design. The multiplexing scheme was validated using a simulation framework that models the PSAPDs and front-end multiplexing circuits to predict the signal-to-noise ratio and flood histogram performance. Two independent experimental setups measured the energy resolution, time resolution, crystal identification ability and count rate both with and without multiplexing. With multiplexing, there was no significant degradation in energy resolution, time resolution and count rate. There was a relative 6.9 ± 1.0% and 9.4 ± 1.0% degradation in the figure of merit that characterizes the crystal identification ability observed in the measured and simulated ceramic-mounted PSAPD module flood histograms, respectively.
Figures
























Similar articles
-
Effects of multiple-interaction photon events in a high-resolution PET system that uses 3-D positioning detectors.Med Phys. 2010 Oct;37(10):5494-508. doi: 10.1118/1.3483262. Med Phys. 2010. PMID: 21089785 Free PMC article.
-
Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography.Phys Med Biol. 2007 Jul 7;52(13):3753-72. doi: 10.1088/0031-9155/52/13/007. Epub 2007 May 29. Phys Med Biol. 2007. PMID: 17664575 Free PMC article.
-
Signal and noise properties of position-sensitive avalanche photodiodes.Phys Med Biol. 2011 Oct 7;56(19):6327-36. doi: 10.1088/0031-9155/56/19/011. Epub 2011 Sep 6. Phys Med Biol. 2011. PMID: 21896961 Free PMC article.
-
A study of the timing properties of position-sensitive avalanche photodiodes.Phys Med Biol. 2009 Sep 7;54(17):5155-72. doi: 10.1088/0031-9155/54/17/006. Epub 2009 Aug 11. Phys Med Biol. 2009. PMID: 19671971 Free PMC article.
-
Bipolar analog signal multiplexing for position-sensitive PET block detectors.Phys Med Biol. 2014 Dec 21;59(24):7835-46. doi: 10.1088/0031-9155/59/24/7835. Phys Med Biol. 2014. PMID: 25419674
Cited by
-
Promising new photon detection concepts for high-resolution clinical and preclinical PET.J Nucl Med. 2012 Feb;53(2):167-70. doi: 10.2967/jnumed.110.084343. J Nucl Med. 2012. PMID: 22302960 Free PMC article. Review.
-
Performance Comparison of Different Readouts for Position-Sensitive Solid-State Photomultiplier Arrays.Biomed Phys Eng Express. 2017 Aug;3(4):045019. doi: 10.1088/2057-1976/aa7c6a. Epub 2017 Jul 19. Biomed Phys Eng Express. 2017. PMID: 29915669 Free PMC article.
-
Convex optimization of coincidence time resolution for a high-resolution PET system.IEEE Trans Med Imaging. 2011 Feb;30(2):391-400. doi: 10.1109/TMI.2010.2080282. Epub 2010 Sep 27. IEEE Trans Med Imaging. 2011. PMID: 20876008 Free PMC article.
-
Thermal regulation of tightly packed solid-state photodetectors in a 1 mm(3) resolution clinical PET system.Med Phys. 2015 Jan;42(1):305-13. doi: 10.1118/1.4903889. Med Phys. 2015. PMID: 25563270 Free PMC article.
-
Characterization of four readout circuits for an MR compatible, preclinical PET detector.Phys Med Biol. 2020 Jun 15;65(12):125008. doi: 10.1088/1361-6560/ab8d76. Phys Med Biol. 2020. PMID: 32340015 Free PMC article.
References
-
- Abreu MC, et al. Design and evaluation of the clear-PEM scanner for positron emission mammography. IEEE Trans. Nucl. Sci. 2006;53:71–77.
-
- Dokhale P, et al. Performancemeasurements of a depth-encoding PET detectormodule based on position-sensitive avalanche photodiode read-out. Phys. Med. Biol. 2004;49:4293–4304. - PubMed
-
- Doshi NK, et al. maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer. IEEE Trans. Nucl. Sci. 2001;48:811–815.
-
- Farrell R. PSAPD device characteristics, private communication. 2009
-
- Freifelder R, Karp JS. Dedicated PET scanners for breast imaging. Phys. Med. Biol. 1997;42:2463–2480. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources