Analytical validation of accelerator mass spectrometry for pharmaceutical development
- PMID: 21083256
- PMCID: PMC3114459
- DOI: 10.4155/bio.10.14
Analytical validation of accelerator mass spectrometry for pharmaceutical development
Abstract
The validation parameters for pharmaceutical analyses were examined for the accelerator mass spectrometry measurement of (14)C/C ratio, independent of chemical separation procedures. The isotope ratio measurement was specific (owing to the (14)C label), stable across samples storage conditions for at least 1 year, linear over four orders of magnitude with an analytical range from 0.1 Modern to at least 2000 Modern (instrument specific). Furthermore, accuracy was excellent (between 1 and 3%), while precision expressed as coefficient of variation was between 1 and 6% determined primarily by radiocarbon content and the time spent analyzing a sample. Sensitivity, expressed as LOD and LLOQ was 1 and 10 attomoles of (14)C, respectively (which can be expressed as compound equivalents) and for a typical small molecule labeled at 10% incorporated with (14)C corresponds to 30 fg equivalents. Accelerator mass spectrometry provides a sensitive, accurate and precise method of measuring drug compounds in biological matrices.
Conflict of interest statement
The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties.
No writing assistance was utilized in the production of this manuscript.
Figures














References
Bibliography
-
- Viswanathan CT, Bansai S, Booth B, et al. Quantitative bioanalytical methods validation and implementation: best practices for chromatographic and ligand binding assays. Pharm. Res. 2007;10:1962–1973. - PubMed
-
- US Pharmacopeia. United States Pharmacopeial Convention, Inc. 1994;23:1982–1984.
-
- ICH International Conference on Harmonisation. Draft guideline on validation of analytical procedures: definitions and terminology. Federal Register. 1995;60:11260.
-
- US FDA. Guideline for submitting samples and analytical data for methods validation. 1987
-
- US FDA. Guidance for industry: Q2B validation of analytical procedures: methodology CDER/CBER. 1996
Websites
-
- GraphPad Software. www.GraphPad.com.
-
- QuantumSoft. www.quansoft.com.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources