BION microstimulators: a case study in the engineering of an electronic implantable medical device
- PMID: 21087890
- DOI: 10.1016/j.medengphy.2010.08.010
BION microstimulators: a case study in the engineering of an electronic implantable medical device
Abstract
The BION (Bionic Neuron) is a single channel implantable neurostimulator of unique design that can be delivered by injection. The development of the BION injectable neurostimulators exemplifies a challenging, but well posed medical design problem addressed with a successful strategy for prioritizing and resolving the biomedical and technological challenges. Though some performance requirements required post-evaluation revision, all fundamental goals were realized. A small number of significant design corrections occurred because the device requirements did not include the full scope of environmental demands. The design has spawned a number of variants optimized for diverse biomedical applications, and its clinical applications continue to evolve. The BION development history demonstrates design successes worth emulating and design pitfalls that may be avoidable for future medical device development teams. This paper serves as an introduction to the BION microstimulator technology and as an analysis of the design process used to develop the early clinical devices.
Copyright © 2010 IPEM. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
BIONic WalkAide for correcting foot drop.IEEE Trans Neural Syst Rehabil Eng. 2005 Jun;13(2):242-6. doi: 10.1109/TNSRE.2005.847385. IEEE Trans Neural Syst Rehabil Eng. 2005. PMID: 16003906
-
A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems.J Peripher Nerv Syst. 2005 Sep;10(3):229-58. doi: 10.1111/j.1085-9489.2005.10303.x. J Peripher Nerv Syst. 2005. PMID: 16221284 Review.
-
BION system for distributed neural prosthetic interfaces.Med Eng Phys. 2001 Jan;23(1):9-18. doi: 10.1016/s1350-4533(01)00011-x. Med Eng Phys. 2001. PMID: 11344003
-
Neural prostheses in clinical applications--trends from precision mechanics towards biomedical microsystems in neurological rehabilitation.Biomed Tech (Berl). 2004 Apr;49(4):72-7. doi: 10.1515/BMT.2004.015. Biomed Tech (Berl). 2004. PMID: 15171585 Review.
-
Improving the drug selection and development process for combination devices.Drug Discov Today. 2007 Oct;12(19-20):844-52. doi: 10.1016/j.drudis.2007.07.020. Epub 2007 Sep 6. Drug Discov Today. 2007. PMID: 17933686 Review.
Cited by
-
Treatment of stress urinary incontinence with a percutaneously implantable wireless microstimulator device (NuStim®) plus pelvic floor muscle exercises: a pilot study.Bladder (San Franc). 2022 Sep 14;9(1):e48. doi: 10.14440/bladder.2022.845. eCollection 2022. Bladder (San Franc). 2022. PMID: 36425078 Free PMC article.
-
Safety and Effectiveness of Nustim System Implantation in Dogs.Bladder (San Franc). 2022 Sep 14;9(1):e49. doi: 10.14440/bladder.2022.844. eCollection 2022. Bladder (San Franc). 2022. PMID: 36425079 Free PMC article.
-
Remote electrical stimulation by means of implanted rectifiers.PLoS One. 2011;6(8):e23456. doi: 10.1371/journal.pone.0023456. Epub 2011 Aug 5. PLoS One. 2011. PMID: 21850274 Free PMC article.
-
A long-lasting wireless stimulator for small mammals.Front Neuroeng. 2013 Oct 11;6:8. doi: 10.3389/fneng.2013.00008. eCollection 2013. Front Neuroeng. 2013. PMID: 24130527 Free PMC article.
-
MiniVStimA: A miniaturized easy to use implantable electrical stimulator for small laboratory animals.PLoS One. 2020 Oct 30;15(10):e0241638. doi: 10.1371/journal.pone.0241638. eCollection 2020. PLoS One. 2020. PMID: 33125415 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources