Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans
- PMID: 21093168
- DOI: 10.1016/j.ijrobp.2010.09.010
Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans
Abstract
Purpose: Fast and reliable comprehensive quality assurance tools are required to validate the safety and accuracy of complex intensity-modulated radiotherapy (IMRT) plans for prostate treatment. In this study, we evaluated the performance of the COMPASS system for both off-line and potential online procedures for the verification of IMRT treatment plans.
Methods and materials: COMPASS has a dedicated beam model and dose engine, it can reconstruct three-dimensional dose distributions on the patient anatomy based on measured fluences using either the MatriXX two-dimensional (2D) array (offline) or a 2D transmission detector (T2D) (online). For benchmarking the COMPASS dose calculation, various dose-volume indices were compared against Monte Carlo-calculated dose distributions for five prostate patient treatment plans. Gamma index evaluation and absolute point dose measurements were also performed in an inhomogeneous pelvis phantom using extended dose range films and ion chamber for five additional treatment plans.
Results: MatriXX-based dose reconstruction showed excellent agreement with the ion chamber (<0.5%, except for one treatment plan, which showed 1.5%), film (∼100% pixels passing gamma criteria 3%/3 mm) and mean dose-volume indices (<2%). The T2D based dose reconstruction showed good agreement as well with ion chamber (<2%), film (∼99% pixels passing gamma criteria 3%/3 mm), and mean dose-volume indices (<5.5%).
Conclusion: The COMPASS system qualifies for routine prostate IMRT pretreatment verification with the MatriXX detector and has the potential for on-line verification of treatment delivery using T2D.
Copyright © 2011 Elsevier Inc. All rights reserved.
Similar articles
-
Volumetric-modulated arc therapy: effective and efficient end-to-end patient-specific quality assurance.Int J Radiat Oncol Biol Phys. 2012 Apr 1;82(5):1567-74. doi: 10.1016/j.ijrobp.2011.01.018. Epub 2011 Apr 4. Int J Radiat Oncol Biol Phys. 2012. PMID: 21470797
-
Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT.Radiother Oncol. 2011 Sep;100(3):446-52. doi: 10.1016/j.radonc.2011.09.007. Epub 2011 Sep 29. Radiother Oncol. 2011. PMID: 21963288
-
3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID.Radiother Oncol. 2007 Feb;82(2):201-7. doi: 10.1016/j.radonc.2006.12.010. Epub 2007 Feb 6. Radiother Oncol. 2007. PMID: 17287039
-
A six-year review of more than 13,000 patient-specific IMRT QA results from 13 different treatment sites.J Appl Clin Med Phys. 2014 Sep 8;15(5):4935. doi: 10.1120/jacmp.v15i5.4935. J Appl Clin Med Phys. 2014. PMID: 25207581 Free PMC article. Review.
-
What is plan quality in radiotherapy? The importance of evaluating dose metrics, complexity, and robustness of treatment plans.Radiother Oncol. 2020 Dec;153:26-33. doi: 10.1016/j.radonc.2020.09.038. Epub 2020 Sep 25. Radiother Oncol. 2020. PMID: 32987045 Review.
Cited by
-
Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file analysis against conventional measurement based IMRT QA.J Appl Clin Med Phys. 2012 Sep 6;13(5):3837. doi: 10.1120/jacmp.v13i5.3837. J Appl Clin Med Phys. 2012. PMID: 22955649 Free PMC article.
-
Commissioning and validation of fluence-based 3D VMAT dose reconstruction system using new transmission detector.Radiol Phys Technol. 2018 Jun;11(2):165-173. doi: 10.1007/s12194-018-0451-8. Epub 2018 Mar 12. Radiol Phys Technol. 2018. PMID: 29532322
-
Assessment of radiobiological metrics applied to patient-specific QA process of VMAT prostate treatments.J Appl Clin Med Phys. 2016 Mar 8;17(2):341-367. doi: 10.1120/jacmp.v17i2.5783. J Appl Clin Med Phys. 2016. PMID: 27074458 Free PMC article.
-
Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS.J Appl Clin Med Phys. 2014 Jan 7;16(1):5128. doi: 10.1120/jacmp.v16i1.5128. J Appl Clin Med Phys. 2014. PMID: 25679152 Free PMC article.
-
Validation of fluence-based 3D IMRT dose reconstruction on a heterogeneous anthropomorphic phantom using Monte Carlo simulation.J Appl Clin Med Phys. 2015 Jan 8;16(1):5199. doi: 10.1120/jacmp.v16i1.5199. J Appl Clin Med Phys. 2015. PMID: 25679177 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical