Comparison of left ventricular functional parameters obtained from three different commercial automated software cardiac quantification program packages and their intraobserver reproducibility
- PMID: 21104169
- DOI: 10.1007/s12149-010-0435-1
Comparison of left ventricular functional parameters obtained from three different commercial automated software cardiac quantification program packages and their intraobserver reproducibility
Abstract
Objective: ECG-gated myocardial perfusion scintigraphy (MPS) can be used to determine several cardiac functional parameters (e.g., left ventricular ejection fraction (LVEF), end-diastolic volume (EDV), and end-systolic volume (ESV)). In this study, we aimed to compare these cardiac functional parameters calculated by the following cardiac quantification programs: Emory Cardiac Toolbox (ECTb), Quantitative Gated SPECT (QGS), and Myometrix. We also evaluated reproducibility of the cardiac programs.
Methods: Fifty-seven patients (27 male, 30 female) at Elazig Research and Training Hospital from 2008 to 2009 were included in this study. In all patients, (99m)Tc-MIBI ECG-Gated (8-bin frame mode) myocardial perfusion scintigraphies were performed. By using 3 different cardiac quantification programs (ECTb, QGS, and Myometrix); LVEF, EDV, and ESV were calculated. The same raw data of MPS images were reprocessed at different time periods, and these 3 parameters were recalculated. LVEF, EDV, and ESV yielded by 3 different programs were compared for interprogram variability assessment, and parameters calculated at two different time periods were compared to evaluate intraprogram reproducibility.
Results: There were statistically significant differences between ECTb, QGS, and Myometrix programs for LVEF, EDV, and ESV (p < 0.001). There was also a statistically significant correlation between LVEF and EDV (p < 0.001, r = 0.546; p < 0.001, r = 0.45, respectively), but no statistically significant correlation was present between the ESV values (p > 0.05, r = 0.09). Statistically significant differences were not found between the values of LVEF, EDV, and ESV obtained from the first and second reconstruction analysis of 3 cardiac quantification programs.
Discussion: Different MPS cardiac software programs give variable (but correlated) LVEF and left ventricular volumetric measures. Those obtained from different cardiac softwares cannot be used interchangeably. Our findings have shown that ECTb, QGS, and Myometrix programs are reproducible, with respect to LVEF, EDV, and ESV.
Similar articles
-
Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT: MRI validation and comparison of the Emory Cardiac Tool Box with QGS and 4D-MSPECT.J Nucl Med. 2005 Aug;46(8):1256-63. J Nucl Med. 2005. PMID: 16085580 Clinical Trial.
-
Comparison of two software in gated myocardial perfusion single photon emission tomography, for the measurement of left ventricular volumes and ejection fraction, in patients with and without perfusion defects.Hell J Nucl Med. 2007 Jan-Apr;10(1):19-23. Hell J Nucl Med. 2007. PMID: 17450245
-
Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET: comparison with cardiac magnetic resonance imaging.Eur J Nucl Med Mol Imaging. 2004 Apr;31(4):482-90. doi: 10.1007/s00259-003-1411-y. Epub 2004 Jan 14. Eur J Nucl Med Mol Imaging. 2004. PMID: 14722679 Clinical Trial.
-
Comparison of software programs for the assessment of left ventricular ejection fraction using 99mTc-tetrofosmin-gated SPECT/CT: correlation with equilibrium radionuclide ventriculography in the Indian population.Nucl Med Commun. 2012 Nov;33(11):1160-8. doi: 10.1097/MNM.0b013e328358cdee. Nucl Med Commun. 2012. PMID: 22955186 Clinical Trial.
-
Are there normal values of phase analysis parameters for left ventricular dyssynchrony in patients with no structural cardiomyopathy?: a systematic review.Nucl Med Commun. 2019 Oct;40(10):980-985. doi: 10.1097/MNM.0000000000001068. Nucl Med Commun. 2019. PMID: 31469810
Cited by
-
Myocardial perfusion and left ventricular quantitative parameters obtained using gated myocardial SPECT: Comparison of three software packages.J Nucl Cardiol. 2018 Jun;25(3):911-924. doi: 10.1007/s12350-016-0730-x. Epub 2016 Nov 21. J Nucl Cardiol. 2018. PMID: 27873167
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources