Intra and interoperator reproducibility of retinal nerve fibre and macular thickness measurements using Cirrus Fourier-domain OCT
- PMID: 21106044
- DOI: 10.1111/j.1755-3768.2010.02045.x
Intra and interoperator reproducibility of retinal nerve fibre and macular thickness measurements using Cirrus Fourier-domain OCT
Abstract
Purpose: To test intrasession, intersession, intervisit and interoperator reproducibility of retinal nerve fibre (RNFL) measurements and retinal thickness in healthy subjects using Cirrus Fourier-domain optical coherence tomography (OCT) (Carl Zeiss Meditec, Inc).
Methods: Seventy-two eyes of 72 healthy subjects were included in the study. All the eyes underwent three 512 × 128 volume cube centred on the fovea and three 360° circular scans centred on the optic disc by one observer. This sequence was redone by another observer on a second visit within a 2-week period. Descriptive statistics, analysis of variance, intraclass correlation coefficients (ICCs) and coefficients of variation (COVs) were calculated for the nine areas corresponding to the Early Treatment Diabetic Retinopathy Study and for quadrants and RNFL clock hr sectors.
Results: Retinal thickness and RNFL measurements were highly reproducible. Mean total retinal thickness was 285.2 ± 15.3 μm by observer 1 and 284.2 ± 12.9 μm by observer 2. Mean COV was 1.2%. Mean RNFL average thicknesses were 96.0 ± 7.7 and 95.7 ± 7.9 μm by observer 1 and 2, respectively. Mean COV was 4.4%. The ICCs ranged from 0.823 to 0.992. Mean differences between both operators were lower than 3 μm, and no significant differences were found.
Conclusions: Retinal and RNFL thickness measurements obtained using Cirrus OCT show good repeatability for healthy eyes and few differences between intra- and interobserver evaluations. It can be considered a valid device for measuring retinal and optic nerve parameters in normal eyes.
© 2010 The Authors. Acta Ophthalmologica © 2010 Acta Ophthalmologica Scandinavica Foundation.
Similar articles
-
Reproducibility of optical coherence tomography measurements in children.Am J Ophthalmol. 2013 Jan;155(1):171-176.e1. doi: 10.1016/j.ajo.2012.06.012. Epub 2012 Sep 8. Am J Ophthalmol. 2013. PMID: 22967864
-
Comparison of three different optical coherence tomography models for total macular thickness measurements in healthy controls.Ophthalmologica. 2009;223(6):352-6. doi: 10.1159/000226600. Epub 2009 Jul 2. Ophthalmologica. 2009. PMID: 19571601 Clinical Trial.
-
Comparison of retinal nerve fiber layer thickness in normal eyes using time-domain and spectral-domain optical coherence tomography.Am J Ophthalmol. 2010 Dec;150(6):807-14. doi: 10.1016/j.ajo.2010.06.024. Am J Ophthalmol. 2010. PMID: 20888550
-
Image quality affects macular and retinal nerve fiber layer thickness measurements on fourier-domain optical coherence tomography.Ophthalmic Surg Lasers Imaging. 2011 May-Jun;42(3):216-21. doi: 10.3928/15428877-20110324-01. Epub 2011 Mar 31. Ophthalmic Surg Lasers Imaging. 2011. PMID: 21449533
-
Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: a variability and diagnostic performance study.Ophthalmology. 2009 Jul;116(7):1257-63, 1263.e1-2. doi: 10.1016/j.ophtha.2009.04.013. Epub 2009 May 22. Ophthalmology. 2009. PMID: 19464061
Cited by
-
Comparison of retinal nerve fiber layer thickness measurements in healthy subjects using fourier and time domain optical coherence tomography.J Ophthalmol. 2012;2012:107053. doi: 10.1155/2012/107053. Epub 2012 May 29. J Ophthalmol. 2012. PMID: 22690327 Free PMC article.
-
Repeatability and reproducibility of retinal nerve fibre layer thickness measurements with the iVue-100 optical coherence tomographer.Afr Health Sci. 2018 Jun;18(2):304-312. doi: 10.4314/ahs.v18i2.15. Afr Health Sci. 2018. PMID: 30602957 Free PMC article.
-
A deep learning approach to predict visual field using optical coherence tomography.PLoS One. 2020 Jul 6;15(7):e0234902. doi: 10.1371/journal.pone.0234902. eCollection 2020. PLoS One. 2020. PMID: 32628672 Free PMC article.
-
Automated Retinal Layer Segmentation Using Spectral Domain Optical Coherence Tomography: Evaluation of Inter-Session Repeatability and Agreement between Devices.PLoS One. 2016 Sep 2;11(9):e0162001. doi: 10.1371/journal.pone.0162001. eCollection 2016. PLoS One. 2016. PMID: 27588683 Free PMC article.
-
Interocular Symmetry and Intermachine Reproducibility of Optic Disc and Macular Parameters Measured by Two Different Models of Optical Coherence Tomography.Clin Ophthalmol. 2024 Aug 26;18:2397-2406. doi: 10.2147/OPTH.S465360. eCollection 2024. Clin Ophthalmol. 2024. PMID: 39221077 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources