Near-infrared fluorescence (NIRF) imaging in breast-conserving surgery: assessing intraoperative techniques in tissue-simulating breast phantoms
- PMID: 21106329
- DOI: 10.1016/j.ejso.2010.10.006
Near-infrared fluorescence (NIRF) imaging in breast-conserving surgery: assessing intraoperative techniques in tissue-simulating breast phantoms
Abstract
Purpose: Breast-conserving surgery (BCS) results in tumour-positive surgical margins in up to 40% of the patients. Therefore, new imaging techniques are needed that support the surgeon with real-time feedback on tumour location and margin status. In this study, the potential of near-infrared fluorescence (NIRF) imaging in BCS for pre- and intraoperative tumour localization, margin status assessment and detection of residual disease was assessed in tissue-simulating breast phantoms.
Methods: Breast-shaped phantoms were produced with optical properties that closely match those of normal breast tissue. Fluorescent tumour-like inclusions containing indocyanine green (ICG) were positioned at predefined locations in the phantoms to allow for simulation of (i) preoperative tumour localization, (ii) real-time NIRF-guided tumour resection, and (iii) intraoperative margin assessment. Optical imaging was performed using a custom-made clinical prototype NIRF intraoperative camera.
Results: Tumour-like inclusions in breast phantoms could be detected up to a depth of 21 mm using a NIRF intraoperative camera system. Real-time NIRF-guided resection of tumour-like inclusions proved feasible. Moreover, intraoperative NIRF imaging reliably detected residual disease in case of inadequate resection.
Conclusion: We evaluated the potential of NIRF imaging applications for BCS. The clinical setting was simulated by exploiting tissue-like breast phantoms with fluorescent tumour-like agarose inclusions. From this evaluation, we conclude that intraoperative NIRF imaging is feasible and may improve BCS by providing the surgeon with imaging information on tumour location, margin status, and presence of residual disease in real-time. Clinical studies are needed to further validate these results.
Copyright © 2010 Elsevier Ltd. All rights reserved.
Similar articles
-
Tissue-simulating phantoms for assessing potential near-infrared fluorescence imaging applications in breast cancer surgery.J Vis Exp. 2014 Sep 19;(91):51776. doi: 10.3791/51776. J Vis Exp. 2014. PMID: 25286185 Free PMC article.
-
Role for intraoperative margin assessment in patients undergoing breast-conserving surgery.Ann Surg Oncol. 2007 Apr;14(4):1458-71. doi: 10.1245/s10434-006-9236-0. Epub 2007 Jan 28. Ann Surg Oncol. 2007. PMID: 17260108
-
Intraoperative radiological margin assessment in breast-conserving surgery.Eur J Surg Oncol. 2014 Apr;40(4):449-53. doi: 10.1016/j.ejso.2014.01.002. Epub 2014 Jan 18. Eur J Surg Oncol. 2014. PMID: 24468296
-
Intraoperative assessment of margins in breast conservative surgery--still in use?J Surg Oncol. 2014 Jul;110(1):15-20. doi: 10.1002/jso.23634. Epub 2014 May 24. J Surg Oncol. 2014. PMID: 24863286 Review.
-
Close/positive margins after breast-conserving therapy: additional resection or no resection?Breast. 2013 Aug;22 Suppl 2:S115-7. doi: 10.1016/j.breast.2013.07.022. Breast. 2013. PMID: 24074771 Review.
Cited by
-
Optical Tissue Phantoms for Quantitative Evaluation of Surgical Imaging Devices.Adv Photonics Res. 2023 Jan;4(1):2200194. doi: 10.1002/adpr.202200194. Epub 2022 Oct 13. Adv Photonics Res. 2023. PMID: 36643020 Free PMC article.
-
Early detection of heterotopic ossification using near-infrared optical imaging reveals dynamic turnover and progression of mineralization following Achilles tenotomy and burn injury.J Orthop Res. 2014 Nov;32(11):1416-23. doi: 10.1002/jor.22697. Epub 2014 Aug 2. J Orthop Res. 2014. PMID: 25087685 Free PMC article.
-
Tissue-simulating phantoms for assessing potential near-infrared fluorescence imaging applications in breast cancer surgery.J Vis Exp. 2014 Sep 19;(91):51776. doi: 10.3791/51776. J Vis Exp. 2014. PMID: 25286185 Free PMC article.
-
Preclinical studies and prospective clinical applications for bacteria-targeted imaging: the future is bright.Clin Transl Imaging. 2016;4:253-264. doi: 10.1007/s40336-016-0190-y. Epub 2016 Jul 16. Clin Transl Imaging. 2016. PMID: 27512688 Free PMC article. Review.
-
Advancing Molecular-Guided Surgery through probe development and testing in a moderate cost evaluation pipeline.Proc SPIE Int Soc Opt Eng. 2015 Mar 4;9311:931112. doi: 10.1117/12.2083224. Proc SPIE Int Soc Opt Eng. 2015. PMID: 25914500 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical