Impact of stent design on the outcome of intervention for carotid bifurcation stenosis
- PMID: 21124276
Impact of stent design on the outcome of intervention for carotid bifurcation stenosis
Abstract
Over the past several years, there has been continued significant interest in refinement of patient selection, devices, procedures and protocols in an effort to optimize the outcome of percutaneous intervention for carotid bifurcation stenosis, including: ongoing National Institutes of Health and manufacturer trials and registries; the further refinement of existing devices and emergence of new platforms to attain distal embolic protection; ongoing study of what really constitutes a high-risk carotid surgery or stenting patient; and attention to device characteristics and patient-device matching. Within the latter area, considerable interest has focused on stent characteristics that have the potential to impact short and long-term outcome when compared with other stent design strategies when studied in large series. The stent in carotid artery intervention occupies a unique role in that after the embolic protection system has been removed, it is the main line of defense (in concert with aggressive dual antiplatelet therapy) from embolic and thromboembolic complications that may arise from the newly remodeled plaque after post-stent angioplasty. In this review, we aim to update the current status of efforts to relate stent design strategy to outcome in intervention for extracranial carotid artery disease with a focus primarily on the function of "free cell area" (typically lower with closed-cell stents and higher with open-cell stents) in analyses of outcome in carotid artery stenting. Also, the potential role of closed-cell vs. open-cell stent selection in other reports related to carotid artery stenting outcome or complications is reviewed. Rigorous studies have examined the issue of free cell area and arrived at disparate conclusions. Randomized data on the impact of free cell area and cell design strategy on carotid intervention are presently lacking. However, we believe sufficient data and rationale exist 1) to warrant ongoing study of the impact of stent design on outcome in carotid intervention; and 2) to make consideration of closed-cell (low free cell area) stent use a reasonable approach to device selection--when patient factors, lesion characteristics, or device availability make doing so possible.
Similar articles
-
Randomized clinical trial of open-cell vs closed-cell stents for carotid stenting and effects of stent design on cerebral embolization.J Vasc Surg. 2011 Nov;54(5):1310-1316.e1; discussion 1316. doi: 10.1016/j.jvs.2011.05.013. Epub 2011 Jul 1. J Vasc Surg. 2011. PMID: 21723064 Clinical Trial.
-
Outcome of carotid artery stenting at 2 years follow-up: comparison of nitinol open cell versus stainless steel closed cell stent design.J Cardiovasc Surg (Torino). 2009 Oct;50(5):669-75. Epub 2009 May 19. J Cardiovasc Surg (Torino). 2009. PMID: 19455089
-
Prevalence and clinical significance of stent fracture and deformation following carotid artery stenting.J Vasc Surg. 2011 Sep;54(3):685-90. doi: 10.1016/j.jvs.2011.03.257. J Vasc Surg. 2011. PMID: 21889703
-
Carotid artery stenting renaissance: is it safe and effective using new materials?J Cardiovasc Surg (Torino). 2016 Dec;57(6):769-783. Epub 2016 Sep 14. J Cardiovasc Surg (Torino). 2016. PMID: 27626627 Review.
-
Will mesh-covered stents help reduce stroke associated with carotid stent angioplasty?Semin Vasc Surg. 2017 Mar;30(1):25-30. doi: 10.1053/j.semvascsurg.2017.04.007. Epub 2017 Apr 27. Semin Vasc Surg. 2017. PMID: 28818256 Review.
Cited by
-
Evaluation and management of plaque protrusion or thrombus following carotid artery stenting.Neurol Med Chir (Tokyo). 2015;55(2):149-54. doi: 10.2176/nmc.oa.2014-0105. Epub 2015 Jan 23. Neurol Med Chir (Tokyo). 2015. PMID: 25746309 Free PMC article.
-
Clinical Outcomes of Second- versus First-Generation Carotid Stents: A Systematic Review and Meta-Analysis.J Clin Med. 2022 Aug 17;11(16):4819. doi: 10.3390/jcm11164819. J Clin Med. 2022. PMID: 36013058 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous