Objective: To investigate the performance of an open MRI system at its conceptual limits by examining excessively obese patients who otherwise could not receive adequate imaging examinations.
Methods: Twenty-six excessively obese patients (BMI ≥ 35, average age 46) where CT, standard MR or ultrasound examinations were not possible or not conclusive were referred to an open MRI system at 1.0 Tesla. Image quality was measured by SNR and CNR with the integrated body coil for obese patients and optimal body coils for a regular weight control group (average BMI 23, average age 30). MRI findings were evaluated by a diagnostic impact matrix.
Results: SNR and CNR were generally lower in obese patients when the integrated body coil was used compared to the normal weight group with ideal body coils e.g.: For cerebral imaging T2W TSE (<5% for white matter, ca. 30% for grey matter) and T1W SE (ca. 15% for white matter, <5% for grey matter), for spinal imaging T2W TSE (ca. 35% for disc and vertebral body) and T1W SE (about 2% for disc, ca. 10% for vertebral body). Relevant new diagnoses impacting patient's therapy were identified in 30% (8/26), the particular medical question of the referring physician could be ruled out as possible reason for the medical condition in 53% (14/26).
Conclusion: In excessively obese patients where CT, standard MR or ultrasound examination is not possible or not conclusive open MRI system have great potential in diagnostic evaluation, offering lower but sufficient image quality to impact therapy.
Similar articles
-
Brain imaging: Comparison of T1W FLAIR BLADE with conventional T1W SE.Magn Reson Imaging. 2017 Apr;37:234-242. doi: 10.1016/j.mri.2016.12.007. Epub 2016 Dec 7. Magn Reson Imaging. 2017. PMID: 27939435
-
Contrast enrichment of spinal cord MR imaging using a ratio of T1-weighted and T2-weighted signals.J Magn Reson Imaging. 2014 Nov;40(5):1199-207. doi: 10.1002/jmri.24456. Epub 2013 Nov 4. J Magn Reson Imaging. 2014. PMID: 24395471
-
High-resolution magnetic resonance imaging of the temporomandibular joint: image quality at 1.5 and 3.0 Tesla in volunteers.Invest Radiol. 2007 Jun;42(6):428-34. doi: 10.1097/01.rli.0000262081.23997.6b. Invest Radiol. 2007. PMID: 17507815
-
Tailored interactive sequences for continuous MR-image-guided freehand biopsies of different organs in an open system at 1.0 tesla (T) - Initial experience.Biomed Tech (Berl). 2017 Nov 27;62(6):557-563. doi: 10.1515/bmt-2015-0163. Biomed Tech (Berl). 2017. PMID: 28076290
-
Diagnostic imaging in obesity.Best Pract Res Clin Endocrinol Metab. 2013 Apr;27(2):261-77. doi: 10.1016/j.beem.2013.02.003. Epub 2013 Mar 5. Best Pract Res Clin Endocrinol Metab. 2013. PMID: 23731887 Review.
Cited by
-
Use of Magnetic Resonance Imaging to Quantify Fat and Steatosis in the Pancreas in Patients after Bariatric Surgery: a Retrospective Study.Obes Surg. 2022 Nov;32(11):3666-3674. doi: 10.1007/s11695-022-06278-4. Epub 2022 Sep 19. Obes Surg. 2022. PMID: 36121606
-
Quantification of epicardial adipose tissue in obese patients using an open-bore MR scanner.Eur Radiol Exp. 2022 May 24;6(1):25. doi: 10.1186/s41747-022-00274-0. Eur Radiol Exp. 2022. PMID: 35606555 Free PMC article.
-
[Percutaneous interventions in an open MR system: technical background and clinical indications].Radiologe. 2013 Nov;53(11):993-1000. doi: 10.1007/s00117-012-2462-5. Radiologe. 2013. PMID: 24170286 German.
-
Potential Pitfalls of Steatopygia on Bone Scintigraphy and the Added Value of Single Photon Emission Computed Tomography and Single Photon Emission Computed Tomography/Computed Tomography.Med Princ Pract. 2023;32(6):343-350. doi: 10.1159/000526372. Epub 2022 Aug 8. Med Princ Pract. 2023. PMID: 35940136 Free PMC article.
-
High-field open versus short-bore magnetic resonance imaging of the spine: a randomized controlled comparison of image quality.PLoS One. 2013 Dec 31;8(12):e83427. doi: 10.1371/journal.pone.0083427. eCollection 2013. PLoS One. 2013. PMID: 24391767 Free PMC article. Clinical Trial.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical