Multicenter user evaluation of ACCU-CHEK® Combo, an integrated system for continuous subcutaneous insulin infusion
- PMID: 21129336
- PMCID: PMC3005051
- DOI: 10.1177/193229681000400615
Multicenter user evaluation of ACCU-CHEK® Combo, an integrated system for continuous subcutaneous insulin infusion
Abstract
Background: The aim of this study was to evaluate a newly developed system for insulin delivery incorporating a multifunctional blood glucose meter and a remotely controlled insulin pump (ACCU-CHEK® Combo system) in established pump users with type 1 diabetes. The technology was assessed both from device performance and subject usability perspectives.
Method: A multicenter, prospective, single group study was carried out in five centers in the Netherlands and four centers in the United Kingdom for more than 6 months. The study was divided into two phases: Phase 1 (4 weeks) for device validation purposes and phase 2 (22 weeks) to observe the impact of the system on metabolic control, patient satisfaction [using the Diabetes Treatment Satisfaction Questionnaire (DTSQ)] and device safety.
Results: Eighty subjects completed the planned study period. There were no unexpected device errors. Treatment satisfaction was high at baseline and further increased to study end (DTSQ change version: sum score, 10.6±7.2; scale score range, -18 to +18, p<0.0001). Hemoglobin A1c improved continuously over time, from 7.9% (±0.9%) to 7.7% (±0.8%) at month 3 (p<0.001) and 7.6% (±0.8%) at month 6 (p<0.0001). The frequency of severe hypoglycemia was 0.08 per patient years. There was no case of ketoacidosis.
Conclusions: The new system was evaluated by experienced continuous subcutaneous insulin infusion users as safe in daily practice and associated with favorable treatment satisfaction and a modest improvement in glycemic control.
© 2010 Diabetes Technology Society.
Figures
References
-
- Weissberg-Benchell J, Antisdel-Lomaglio J, Seshadri R. Insulin pump therapy: a meta-analysis. Diabetes Care. 2003;26(4):1079–1087. - PubMed
-
- Barnard K, Lloyd C, Skinner T. Systematic literature review: quality of life associated with insulin pump use in type 1 diabetes. Diabet Med. 2007;24(6):607–617. - PubMed
-
- Speight J, Barendse SM. FDA guidance on patient reported outcomes. BMJ. 2010;340(7761):c2921. - PubMed
-
- Todres L, Kerr D, Keen S. Continuous subcutaneous insulin infusion in type 1 diabetes: patient experiences of ‘living with a machine’. Diabet Med. 2010;27(10):1201–1204. - PubMed
-
- Bradley C, Lewis K. Measures of psychological well-being and treatment satisfaction developed from the responses of people with tablet-treated diabetes. Diabet Med. 1990;7(5):445–451. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical