Reversed-phase liquid chromatography/electrospray ionization/mass spectrometry with isotope dilution for the analysis of nitrate and nitrite in water

Affiliations

Reversed-phase liquid chromatography/electrospray ionization/mass spectrometry with isotope dilution for the analysis of nitrate and nitrite in water

Yongtao Li et al. J Chromatogr A. .

Abstract

A new method was developed for the analysis of nitrate and nitrite in a variety of water matrices by using reversed-phase liquid chromatography/electrospray ionization/mass spectrometry in the negative ion mode. For this direct analysis method, nitrate and nitrite anions were well separated under the optimized LC conditions, detected by monitoring m/z 62 and m/z 46 ions, and quantitated by using an isotope dilution technique that utilized the isotopically labeled analogs. The method sensitivity, accuracy, and precision were investigated, along with matrix effects resulting from common inorganic matrix anions. The isotope dilution technique, along with sample pretreatment using barium, silver, and hydrogen cartridges, effectively compensated for the ionization suppression caused by the major water matrix anions, including chloride, sulfate, phosphate, and carbonate. The method detection limits, based on seven reagent water replicates fortified at 0.01 mg N/L nitrate and 0.1 mg N/L nitrite, were 0.001 mg N/L for nitrate and 0.012-0.014 mg N/L for nitrite. The mean recoveries from the replicate fortified reagent water and lab water samples containing the major water matrix anions, were 92-103% for nitrate with an imprecision (relative standard deviation, RSD) of 0.4-2.1% and 92-110% for nitrite with an RSD of 1.1-4.4%. For the analysis of nitrate and nitrite in drinking water, surface water, and groundwater samples, the obtained results were generally consistent with those obtained from the reference methods. The mean recoveries from the replicate matrix spikes were 92-123% for nitrate with an RSD of 0.6-7.7% and 105-113% for nitrite with an RSD of 0.3-1.8%.