Imageless navigation system does not improve component rotational alignment in total knee arthroplasty
- PMID: 21176919
- DOI: 10.1016/j.jss.2010.05.006
Imageless navigation system does not improve component rotational alignment in total knee arthroplasty
Abstract
Background: The aim of computer-assisted surgery is to improve accuracy and limit the range of surgical variability. However, a worldwide debate exists regarding the importance and usefulness of computer-assisted navigation for total knee arthroplasty (TKA). The main purpose of this study is to summarize and compare the radiographic outcomes of TKA performed using imageless computer-assisted navigation compared with conventional techniques.
Materials and methods: An electronic search of PubMed, EMBASE, Web of Science, and Cochrane library databases was made, in addition to manual search of major orthopedic journals. A meta-analysis of 29 quasi-randomized/randomized controlled trials (quasi-RCTs/RCTs) and 11 prospective comparative studies was conducted through a random effects model. Additional a priori sources of clinical heterogeneity were evaluated by subgroup analysis with regard to radiographic methods.
Results: When the outlier cut-off value of lower limb axis was defined as ±2° or ±3° from the neutral, the postoperative full-length radiographs demonstrated that the risk ratio was 0.54 or 0.39, respectively, which were in favor of the navigated group. When the cut-off value used for the alignment in the coronal and sagittal plane was 2° or 3°, imageless navigation significantly reduced the outlier rate of the femoral and tibial components compared with the conventional group. Notably, computed tomography scans demonstrated no statistically significant differences between the two groups regarding the outliers in the rotational alignment of the femoral and tibial components; however, there was strong statistical heterogeneity.
Conclusions: Our results indicated that imageless computer-assisted navigation systems improve lower limb axis and component orientation in the coronal and sagittal planes, but not the rotational alignment in TKA. Further multiple-center clinical trials with long-term follow-up are needed to determine differences in the clinical and functional outcomes of knee arthroplasties performed using computer-assisted techniques.
Copyright © 2011 Elsevier Inc. All rights reserved.
Similar articles
-
Comparison of robot-assisted and conventional total knee arthroplasty: a controlled cadaver study using multiparameter quantitative three-dimensional CT assessment of alignment.Comput Aided Surg. 2012;17(2):86-95. doi: 10.3109/10929088.2012.654408. Comput Aided Surg. 2012. PMID: 22348661
-
[Rotational alignment of femoral component with computed-assisted surgery (CAS) during total knee arthroplasty].Rev Chir Orthop Reparatrice Appar Mot. 2008 Oct;94(6):580-4. doi: 10.1016/j.rco.2008.03.038. Epub 2008 May 27. Rev Chir Orthop Reparatrice Appar Mot. 2008. PMID: 18929753 French.
-
Rotational alignment in total knee arthroplasty: nonimage-based navigation system versus conventional technique.Chin Med J (Engl). 2012 Jan;125(2):236-43. Chin Med J (Engl). 2012. PMID: 22340552
-
Little clinical advantage of computer-assisted navigation over conventional instrumentation in primary total knee arthroplasty at early follow-up.Knee. 2012 Aug;19(4):237-45. doi: 10.1016/j.knee.2011.10.001. Epub 2011 Nov 29. Knee. 2012. PMID: 22130355 Review.
-
No differences in mid- to long-term outcomes of computer-assisted navigation versus conventional total knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2020 Oct;28(10):3183-3192. doi: 10.1007/s00167-019-05808-5. Epub 2019 Nov 29. Knee Surg Sports Traumatol Arthrosc. 2020. PMID: 31784782 Review.
Cited by
-
Computer-assisted navigation for the intraoperative assessment of lower limb alignment in high tibial osteotomy can avoid outliers compared with the conventional technique.Knee Surg Sports Traumatol Arthrosc. 2013 Jan;21(1):181-8. doi: 10.1007/s00167-012-2088-1. Epub 2012 Jul 4. Knee Surg Sports Traumatol Arthrosc. 2013. PMID: 22760216
-
Influence of intentional femoral component flexion in navigated TKA on gap balance and sagittal anatomy.Knee Surg Sports Traumatol Arthrosc. 2014 Mar;22(3):687-93. doi: 10.1007/s00167-013-2731-5. Epub 2013 Oct 26. Knee Surg Sports Traumatol Arthrosc. 2014. PMID: 24158449
-
No difference in long-term functional outcomes or survivorship after total knee arthroplasty with or without computer navigation: a 17-year survivorship analysis.Knee Surg Relat Res. 2021 Sep 8;33(1):30. doi: 10.1186/s43019-021-00114-2. Knee Surg Relat Res. 2021. PMID: 34496974 Free PMC article.
-
[Total knee arthroplasty with the use of patient specific instruments. The VISIONAIRE system].Orthopade. 2016 Apr;45(4):286-93. doi: 10.1007/s00132-016-3239-8. Orthopade. 2016. PMID: 26952362 German.
-
Differences in component and limb alignment between computer-assisted and conventional surgery total knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2014 Dec;22(12):2954-61. doi: 10.1007/s00167-014-3331-8. Epub 2014 Oct 21. Knee Surg Sports Traumatol Arthrosc. 2014. PMID: 25331655
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical