Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry
- PMID: 21196692
- DOI: 10.1159/000322109
Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry
Abstract
Background: Accurate and comfortable methods are needed to determine resting energy expenditure (REE) in older people who are characterized by a lowered metabolic rate. The portable SenseWear® armband (SWA) body monitor, worn on the right upper arm, can easily be used by this age group in an ambulatory manner.
Objective: The purpose of this study was to evaluate the reliability and accuracy of the SWA armband in determining REE in healthy, normal-weight older people.
Methods: Participants were 49 older, community-dwelling volunteers aged 60-87 years. Reliability was tested in all subjects. Accuracy was analyzed in 32 subjects by comparing REE estimated by SWA against REE measured by indirect calorimetry as a criterion method. Data were simultaneously and continuously recorded for 20 min in the morning. In the same subjects, REE determined by SWA from night recording was evaluated for accuracy against indirect calorimetry. The agreement between methods was assessed by the Bland-Altman procedure.
Results: (1) REE measured by SWA was reliable when comparing 2 consecutive measurements (typical error: 1.9%). (2) REE calculated from morning recording (1,543 ± 181 kcal/24 h) was higher than REE from night recording (1,564 ± 192 kcal/24 h; p = 0.018). (3) Compared with REE by indirect calorimetry (1,377 ± 228 kcal/24 h) for accuracy, the SWA overestimated mean REE in the older people by 12 and 14%, respectively, for morning and night recording (each p < 0.001).
Conclusions: The SWA provides a reliable estimate of REE in healthy older subjects and has the advantage of easy handling. The 20-min recording time, which was recommended by the manufacturer, can be applied. However, the SWA overestimates REE in this group, possibly due to age-related changes in skin conductance and thermoregulation, both being measured by sensors of the SWA armband. This requires improving the SWA by developing better fitting algorithms for predicting REE in older people.
Copyright © 2011 S. Karger AG, Basel.
Similar articles
-
Evaluation of a multisensor armband in estimating energy expenditure in obese individuals.Obesity (Silver Spring). 2006 Dec;14(12):2217-23. doi: 10.1038/oby.2006.260. Obesity (Silver Spring). 2006. PMID: 17189549
-
Accuracy of SenseWear Pro2 Armband to predict resting energy expenditure in childhood obesity.Obesity (Silver Spring). 2013 Dec;21(12):2465-70. doi: 10.1002/oby.20427. Epub 2013 Jul 2. Obesity (Silver Spring). 2013. PMID: 23512749
-
Poor agreement between a portable armband and indirect calorimetry in the assessment of resting energy expenditure.Clin Nutr. 2008 Apr;27(2):307-10. doi: 10.1016/j.clnu.2007.11.005. Epub 2008 Feb 13. Clin Nutr. 2008. PMID: 18276043
-
Resting energy expenditure (REE) in an old-old population: implications for metabolic stress.Exp Gerontol. 2014 Nov;59:47-50. doi: 10.1016/j.exger.2014.06.009. Epub 2014 Jun 11. Exp Gerontol. 2014. PMID: 24929009 Review.
-
Energy requirements in frail elderly people: a review of the literature.Clin Nutr. 2007 Feb;26(1):16-24. doi: 10.1016/j.clnu.2006.08.003. Epub 2006 Oct 10. Clin Nutr. 2007. PMID: 17034905 Review.
Cited by
-
Validity of activity monitors in health and chronic disease: a systematic review.Int J Behav Nutr Phys Act. 2012 Jul 9;9:84. doi: 10.1186/1479-5868-9-84. Int J Behav Nutr Phys Act. 2012. PMID: 22776399 Free PMC article.
-
Evaluation of a non-invasive multisensor accelerometer for calculating energy expenditure in ventilated intensive care patients compared to indirect calorimetry and predictive equations.J Clin Monit Comput. 2017 Oct;31(5):1009-1017. doi: 10.1007/s10877-016-9934-5. Epub 2016 Sep 14. J Clin Monit Comput. 2017. PMID: 27628058
-
Low validity of the Sensewear Pro3 activity monitor compared to indirect calorimetry during simulated free living in patients with osteoarthritis of the hip.BMC Musculoskelet Disord. 2014 Feb 19;15:43. doi: 10.1186/1471-2474-15-43. BMC Musculoskelet Disord. 2014. PMID: 24552503 Free PMC article.
-
Energy expenditure evaluation in humans and non-human primates by SenseWear Armband. Validation of energy expenditure evaluation by SenseWear Armband by direct comparison with indirect calorimetry.PLoS One. 2013 Sep 19;8(9):e73651. doi: 10.1371/journal.pone.0073651. eCollection 2013. PLoS One. 2013. PMID: 24069218 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical