Comparison of the QIAsymphony automated nucleic acid extraction and PCR setup platforms with NucliSens easyMAG and Corbett CAS-1200 liquid handling station for the detection of enteric pathogens in fecal samples
- PMID: 21199666
- DOI: 10.1016/j.mimet.2010.12.024
Comparison of the QIAsymphony automated nucleic acid extraction and PCR setup platforms with NucliSens easyMAG and Corbett CAS-1200 liquid handling station for the detection of enteric pathogens in fecal samples
Abstract
In this study we compared the QIAsymphony Sample Preparation and Assay Setup modules (Qiagen), which provide automated nucleic acid extraction and PCR setup, to the NucliSens easyMAG (bioMérieux) and CAS-1200 liquid handling station (Corbett) for a molecular screening approach of enteric pathogens in fecal samples using multiplex real-time PCR. The relative DNA recovery of both platforms, within- and between-run reproducibility and a prospective study, including 510 clinical fecal samples, were performed. The results demonstrated that the QIAsymphony Sample Preparation and Assay Setup modules were highly reproducible and achieve equal performance, quantitative and qualitative, when compared with the NucliSens easyMAG and CAS-1200 systems for the molecular screening analysis of enteric pathogens in fecal samples.
Copyright © 2010 Elsevier B.V. All rights reserved.
Similar articles
-
Comparative evaluation of the QIAGEN QIAsymphony® SP system and bioMérieux NucliSens easyMAG automated extraction platforms in a clinical virology laboratory.J Clin Virol. 2011 Dec;52(4):339-43. doi: 10.1016/j.jcv.2011.08.016. Epub 2011 Oct 20. J Clin Virol. 2011. PMID: 22014618
-
Evaluation of a shortened QIAsymphony DNA extraction protocol for stool samples using a multiplex real-time PCR for the detection of enteric pathogens.J Microbiol Methods. 2011 Feb;84(2):365-7. doi: 10.1016/j.mimet.2010.12.023. Epub 2010 Dec 31. J Microbiol Methods. 2011. PMID: 21199665
-
Evaluation of NucliSens easyMAG for automated nucleic acid extraction from various clinical specimens.J Clin Microbiol. 2007 Feb;45(2):421-5. doi: 10.1128/JCM.00894-06. Epub 2006 Dec 13. J Clin Microbiol. 2007. PMID: 17166966 Free PMC article.
-
Comparative evaluation of two automated systems for nucleic acid extraction of BK virus: NucliSens easyMAG versus BioRobot MDx.J Virol Methods. 2009 Dec;162(1-2):208-12. doi: 10.1016/j.jviromet.2009.08.007. Epub 2009 Aug 25. J Virol Methods. 2009. PMID: 19712698
-
Extraction of viral nucleic acids: comparison of five automated nucleic acid extraction platforms.J Clin Virol. 2012 Jul;54(3):255-9. doi: 10.1016/j.jcv.2012.03.008. Epub 2012 Apr 13. J Clin Virol. 2012. PMID: 22503856
Cited by
-
Comparative evaluation of the QIAsymphony RGQ system with the easyMAG/R-gene combination for the quantitation of cytomegalovirus DNA load in whole blood.Virol J. 2012 Oct 9;9:231. doi: 10.1186/1743-422X-9-231. Virol J. 2012. PMID: 23046712 Free PMC article.
-
Comparison of DNA Extraction Methods and Real-Time PCR Assays for the Detection of Blastocystis sp. in Stool Specimens.Microorganisms. 2020 Nov 11;8(11):1768. doi: 10.3390/microorganisms8111768. Microorganisms. 2020. PMID: 33187097 Free PMC article.
-
Molecular testing for clinical diagnosis and epidemiological investigations of intestinal parasitic infections.Clin Microbiol Rev. 2014 Apr;27(2):371-418. doi: 10.1128/CMR.00122-13. Clin Microbiol Rev. 2014. PMID: 24696439 Free PMC article. Review.
-
Comparison of eMAG™ versus NucliSENS® EasyMAG® performance on clinical specimens.J Clin Virol. 2017 Mar;88:52-57. doi: 10.1016/j.jcv.2017.01.004. Epub 2017 Jan 29. J Clin Virol. 2017. PMID: 28160729 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical