In vivo dosimetry with diodes in a radiotherapy department in Pakistan
- PMID: 21199828
- DOI: 10.1093/rpd/ncq566
In vivo dosimetry with diodes in a radiotherapy department in Pakistan
Abstract
The International Commission of Radiological Units (ICRU) sets a tolerance of ±5 % on dose delivery, with more recent data limiting the overall tolerances to ±3 %. One of the best methods for accurate dose delivery and quality check is in vivo dosimetry, while radiotherapy is performed. The present study was carried out to test the applicability of diodes for performing in vivo entrance dose measurements in external photon beam radiotherapy for pelvic tumours and its implementation as quality assurance tool in radiotherapy. During November 2007 to December 2009, in 300 patients who received pelvic radiotherapy on a multileaf-collimator-assisted linear accelerator, the central axis dose was measured by in vivo dosimetry by p-Si diodes. Entrance dose measurements were taken by diodes and were compared with the prescribed dose. Totally 1000 calculations were performed. The mean and standard deviation between measured and prescribed dose was 1.26 ± 2.8 %. In 938 measurements (93.8 %), the deviation was <5 % (1.36 ± 2.9%); in 62 measurements (6.2 %) the mean deviation was >5 % (5.51 ± 2.3 %). Larger variations were seen in lateral and oblique fields more than anteroposterior fields. For larger deviations, patients and diode positional errors were found to be the common factors alone or in combination with other factors. After additional corrections, repeated measurements were achieved within tolerance levels. This study showed that diode-detector-based in vivo dosimetry was simple, cost-effective, provides quick results and can serve as a useful quality assurance tool in radiotherapy. The data acquired in the present study can be used for evaluating output calibration of therapy machine, precision of calculations, effectiveness of treatment plan and patient setup.
Similar articles
-
Implementation of in vivo dosimetry with Isorad semiconductor diodes in radiotherapy treatments of the pelvis.Med Dosim. 2008 Spring;33(1):14-21. doi: 10.1016/j.meddos.2007.04.004. Med Dosim. 2008. PMID: 18262118 Clinical Trial.
-
Midline dose verification with diode in vivo dosimetry for external photon therapy of head and neck and pelvis cancers during initial large-field treatments.Med Dosim. 2010 Winter;35(4):304-11. doi: 10.1016/j.meddos.2010.03.007. Epub 2010 Jun 2. Med Dosim. 2010. PMID: 20627683
-
[Results of quality controls for individual doses to patients in radiation oncology].Strahlenther Onkol. 2008 Oct;184(10):505-9. doi: 10.1007/s00066-008-1866-7. Epub 2008 Oct 1. Strahlenther Onkol. 2008. PMID: 19016038 German.
-
[Verification of the dose delivered to the patient by means of TLD, SC, PID. What future?].Cancer Radiother. 2003 Nov;7 Suppl 1:78s-84s. Cancer Radiother. 2003. PMID: 15124548 Review. French.
-
A literature review of electronic portal imaging for radiotherapy dosimetry.Radiother Oncol. 2008 Sep;88(3):289-309. doi: 10.1016/j.radonc.2008.07.008. Epub 2008 Aug 14. Radiother Oncol. 2008. PMID: 18706727 Review.
Cited by
-
Evaluation of target dose based on water-equivalent thickness in external beam radiotherapy.J Med Phys. 2013 Jan;38(1):44-51. doi: 10.4103/0971-6203.106605. J Med Phys. 2013. PMID: 23532059 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources