Outcome analysis in 3,160 implantations of radiologically guided placements of totally implantable central venous port systems
- PMID: 21207035
- DOI: 10.1007/s00330-010-2045-7
Outcome analysis in 3,160 implantations of radiologically guided placements of totally implantable central venous port systems
Abstract
Objectives: In this retrospective study the success and complication rates after radiologically guided port catheter implantation were evaluated.
Methods: Between 2000 and 2008, 3,160 port catheter systems were implanted in our interventional suite. All interventions were imaging guided. The puncture of the preferably right internal jugular vein (IJV) was ultrasound-assisted and the catheter tip position was controlled with fluoroscopy. Catheter indwelling time and rates of periprocedural, early and late complications were evaluated.
Results: 922,599 catheter days (mean, 292 days; range, 0-2,704 days) were documented. The implantation was successful in 3,153 (99.8%) cases. A total of 374 (11.8%; 0.41/1,000 catheter days) adverse events were recorded. Of these, 42 (1.33%) were periprocedural complications. 86 (3.3%; 0.09/1,000 catheter days) early and 246 (9.4%; 0.27/1,000 catheter days) late onset complications occurred after port implantation. The most common complications were blood stream infection (n = 134; 5.1%; 0.15/1,000 catheter days), catheter-induced venous thrombosis (n = 97; 3.7%; 0.11/1,000 catheter days) and catheter migration (n = 34; 1.3%; 0.04/1,000 catheter days). A total of 193 (6.1%) port explantations were required.
Conclusion: Ultrasound guided port implantation via the IJV results in low periprocedural complication rates.
Comment in
-
Letter to the Editor re: "Outcome analysis in 3160 implantations of radiologically guided placements of totally implantable central venous port systems": observations about right internal jugular access.Eur Radiol. 2011 Oct;21(10):2166-7; author reply 2168. doi: 10.1007/s00330-011-2163-x. Epub 2011 May 28. Eur Radiol. 2011. PMID: 21626361 No abstract available.
Similar articles
-
Letter to the Editor re: "Outcome analysis in 3160 implantations of radiologically guided placements of totally implantable central venous port systems": observations about right internal jugular access.Eur Radiol. 2011 Oct;21(10):2166-7; author reply 2168. doi: 10.1007/s00330-011-2163-x. Epub 2011 May 28. Eur Radiol. 2011. PMID: 21626361 No abstract available.
-
Combined ultrasound and fluoroscopy guided port catheter implantation--high success and low complication rate.Eur J Radiol. 2009 Mar;69(3):517-22. doi: 10.1016/j.ejrad.2007.10.018. Eur J Radiol. 2009. PMID: 19340958
-
Ultrasound and fluoroscopy-guided placement of central venous ports via internal jugular vein: retrospective analysis of 1254 port implantations at a single center.Korean J Radiol. 2012 May-Jun;13(3):314-23. doi: 10.3348/kjr.2012.13.3.314. Epub 2012 Apr 17. Korean J Radiol. 2012. PMID: 22563269 Free PMC article.
-
Evaluation of radiologically implanted central venous port systems explanted due to complications.J Vasc Access. 2011 Oct-Dec;12(4):306-12. doi: 10.5301/JVA.2011.7739. J Vasc Access. 2011. PMID: 21534232
-
[Percutaneous interventional radiologic implantation of intravenous port-catheter systems].Radiologe. 1999 Sep;39(9):777-82. doi: 10.1007/s001170050575. Radiologe. 1999. PMID: 10525636 Review. German.
Cited by
-
A Monocentric Analysis of Implantable Ports in Cancer Treatment: Five-Year Efficacy and Safety Evaluation.Cancers (Basel). 2024 Aug 9;16(16):2802. doi: 10.3390/cancers16162802. Cancers (Basel). 2024. PMID: 39199575 Free PMC article.
-
[Superior vena cava occlusion : Radiological placement of a central venous port system via femoral vein access].Radiologe. 2012 May;52(5):455-8. doi: 10.1007/s00117-012-2317-0. Radiologe. 2012. PMID: 22552449 German.
-
Chemoport-A Savior in Children Who Require Chronic Venous Access: An Observational Study.Vasc Specialist Int. 2019 Sep;35(3):145-151. doi: 10.5758/vsi.2019.35.3.145. Epub 2019 Sep 30. Vasc Specialist Int. 2019. PMID: 31620400 Free PMC article.
-
Letter to the Editor re: "Outcome analysis in 3160 implantations of radiologically guided placements of totally implantable central venous port systems": observations about right internal jugular access.Eur Radiol. 2011 Oct;21(10):2166-7; author reply 2168. doi: 10.1007/s00330-011-2163-x. Epub 2011 May 28. Eur Radiol. 2011. PMID: 21626361 No abstract available.
-
Femoral placement of totally implantable venous power ports as an alternative implantation site for patients with central vein occlusions.Support Care Cancer. 2014 Feb;22(2):383-7. doi: 10.1007/s00520-013-1984-3. Epub 2013 Sep 24. Support Care Cancer. 2014. PMID: 24061782
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical