Comparison of everolimus-eluting and paclitaxel-eluting coronary stents in patients undergoing multilesion and multivessel intervention: the SPIRIT III (A Clinical Evaluation of the Investigational Device XIENCE V Everolimus Eluting Coronary Stent System [EECSS] in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) and SPIRIT IV (Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) randomized trials
- PMID: 21232716
- DOI: 10.1016/j.jcin.2010.09.014
Comparison of everolimus-eluting and paclitaxel-eluting coronary stents in patients undergoing multilesion and multivessel intervention: the SPIRIT III (A Clinical Evaluation of the Investigational Device XIENCE V Everolimus Eluting Coronary Stent System [EECSS] in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) and SPIRIT IV (Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) randomized trials
Abstract
Objectives: We evaluated outcomes following XIENCE V everolimus-eluting stent (EES) compared with the Taxus Express(2) paclitaxel-eluting stent (PES) in patients undergoing multilesion and multivessel intervention.
Background: The optimal revascularization strategy for patients with multivessel disease is unknown.
Methods: The SPIRIT III (A Clinical Evaluation of the Investigational Device XIENCE V Everolimus Eluting Coronary Stent System [EECSS] in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) (n = 1,002) and SPIRIT IV (Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) (n = 3,690) trials enrolled patients with de novo lesions ≤ 28 mm in length and reference vessel diameter of 2.5 to 3.75 mm. The SPIRIT III trial enrolled patients with a single lesion in 1 or 2 coronary arteries, and the SPIRIT IV trial enrolled patients with up to 2 lesions in 3 different vessels (maximum 2 lesions per vessel). In both trials, patients were randomized 2:1 to EES vs. PES. Clinical outcomes to 1 year were analyzed in patients with single (n = 3,823) versus multiple (n = 765) treated vessels, and in those with single (n = 3,536) versus multiple (n = 1,052) treated lesions.
Results: Among patients with multivessel disease, EES compared with PES resulted in reduced rates of target vessel myocardial infarction (2.2% vs. 6.1%, p = 0.007) and ischemia-driven target lesion revascularization (4.2% vs. 8.0%, p = 0.04). Among patients undergoing multilesion stenting, EES compared with PES resulted in reduced rates of target vessel myocardial infarction (2.1% vs. 5.4%, p = 0.008) and ischemia-driven target lesion revascularization (3.7% vs. 7.4%, p = 0.01). The absolute benefits of EES versus PES in patients undergoing multivessel or multilesion intervention were greater than in those undergoing single-lesion, single-vessel intervention.
Conclusions: The EES compared with PES provided significant improvements in clinical safety and efficacy outcomes. The absolute benefit provided by EES versus PES appears to be proportional to the complexity of coronary disease.
Trial registration: ClinicalTrials.gov NCT00180479 NCT00307047.
Copyright © 2010 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Clinical follow-up 3 years after everolimus- and paclitaxel-eluting stents: a pooled analysis from the SPIRIT II (A Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions) and SPIRIT III (A Clinical Evaluation of the Investigational Device XIENCE V Everolimus Eluting Coronary Stent System [EECSS] in the Treatment of Subjects With De Novo Native Coronary Artery Lesions) randomized trials.JACC Cardiovasc Interv. 2010 Dec;3(12):1220-8. doi: 10.1016/j.jcin.2010.07.017. JACC Cardiovasc Interv. 2010. PMID: 21232715
-
Comparison of everolimus- and paclitaxel-eluting stents in patients with acute and stable coronary syndromes: pooled results from the SPIRIT (A Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System) and COMPARE (A Trial of Everolimus-Eluting Stents and Paclitaxel-Eluting Stents for Coronary Revascularization in Daily Practice) Trials.JACC Cardiovasc Interv. 2011 Oct;4(10):1104-15. doi: 10.1016/j.jcin.2011.06.018. JACC Cardiovasc Interv. 2011. PMID: 22017936 Clinical Trial.
-
Meta-analysis of everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease: final 3-year results of the SPIRIT clinical trials program (Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions).JACC Cardiovasc Interv. 2013 Sep;6(9):914-22. doi: 10.1016/j.jcin.2013.05.005. JACC Cardiovasc Interv. 2013. PMID: 24050859
-
Percutaneous coronary intervention with everolimus-eluting stents (Xience V): systematic review and direct-indirect comparison meta-analyses with paclitaxel-eluting stents (Taxus) and sirolimus-eluting stents (Cypher).Minerva Cardioangiol. 2008 Feb;56(1):55-65. Minerva Cardioangiol. 2008. PMID: 18432169
-
The paclitaxel (TAXUS)-eluting stent: a review of its use in the management of de novo coronary artery lesions.Am J Cardiovasc Drugs. 2004;4(4):257-68. doi: 10.2165/00129784-200404040-00006. Am J Cardiovasc Drugs. 2004. PMID: 15285700 Review.
Cited by
-
Complex vs. non-complex percutaneous coronary intervention with newer-generation drug-eluting stents: an analysis from the randomized BIOFLOW trials.Clin Res Cardiol. 2022 Jul;111(7):795-805. doi: 10.1007/s00392-022-01994-4. Epub 2022 Feb 25. Clin Res Cardiol. 2022. PMID: 35212802
-
Comparison of neointimal coverage between ultrathin biodegradable polymer-coated sirolimus-eluting stents and durable polymer-coated everolimus-eluting stents: 6 months optical coherence tomography follow-up from the TAXCO study.Catheter Cardiovasc Interv. 2021 Feb 15;97(3):423-430. doi: 10.1002/ccd.28833. Epub 2020 Apr 3. Catheter Cardiovasc Interv. 2021. PMID: 32243050 Free PMC article.
-
Comparison of first- and second-generation drug-eluting stent efficacies for treating left main and/or three-vessel disease: a propensity matched study.Heart Vessels. 2016 Dec;31(12):1930-1942. doi: 10.1007/s00380-016-0824-z. Epub 2016 Mar 2. Heart Vessels. 2016. PMID: 26936450
-
Everolimus-eluting bioresorbable scaffolds for treatment of coronary artery disease in patients with diabetes mellitus: the midterm follow-up of the prospective ABSORB DM Benelux study.Cardiovasc Diabetol. 2019 Mar 9;18(1):25. doi: 10.1186/s12933-019-0827-z. Cardiovasc Diabetol. 2019. PMID: 30851731 Free PMC article.
-
Efficacy of non-compliant balloon post-dilation in optimization of contemporary stents: A digital stent enhancement study.Int J Cardiol Heart Vessel. 2014 Mar 19;3:43-48. doi: 10.1016/j.ijchv.2014.03.006. eCollection 2014 Jun. Int J Cardiol Heart Vessel. 2014. PMID: 29450169 Free PMC article.
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Medical