Comparative Study
. 2011 Oct;95(10):1410-4.
doi: 10.1136/bjo.2009.175869. Epub 2011 Jan 17.

Single versus sequential testing with scanning peripheral anterior chamber depth analyser, IOLMaster and anterior segment optical coherence tomography for the detection of narrow angles

Affiliations
Comparative Study

Single versus sequential testing with scanning peripheral anterior chamber depth analyser, IOLMaster and anterior segment optical coherence tomography for the detection of narrow angles

Dolly S Chang et al. Br J Ophthalmol. 2011 Oct.

Abstract

Background: Scanning peripheral anterior chamber depth analyser (SPAC), IOLMaster and anterior segment optical coherence tomography (AS-OCT) are devices that can screen non-invasively for narrow angles (NAs) and may reduce the need for gonioscopy, especially where the availability of ophthalmologists is limited. When used individually to detect NAs, neither SPAC, AS-OCT nor IOLMaster were able to achieve both high sensitivity and specificity when using gonioscopy as a gold standard.

Aims: To assess whether sequential testing improves discrimination of NAs over single testing.

Methods: In a prospective community-based cross-sectional study in Singapore, 2047 phakic right eyes underwent SPAC, IOLMaster and AS-OCT. NA eyes were defined as present when Shaffer scores were 0 or 1 in ≥ 2 quadrants on gonioscopy. Sensitivities and specificities of single and sequential testing were compared.

Results: At a fixed specificity of 95%, sensitivities in detecting NA eyes were 57.8% (95% CI 52.7% to 62.7%) for SPAC, 43.3% (95% CI 38.3% to 48.3%) for IOLMaster and 52.1% (95% CI 46.4% to 57.7%) for AS-OCT. After sequential testing using SPAC followed by AS-OCT, the sensitivity rose to 70.3% (95% CI 65.5% to 74.8%) while specificity remained at 94.3% (95% CI 93.1% to 95.4%), with sensitivity significantly higher than single testing alone (p<0.001).

Conclusion: Sequential testing significantly improves the accuracy of detecting NAs and may be useful in screening settings.