Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 Tesla MRI
- PMID: 21274985
- DOI: 10.1002/jmri.22463
Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 Tesla MRI
Abstract
Purpose: To assess the risk of RF-induced heating in pacemaker-attached and abandoned leads using in vitro temperature measurements at 1.5 Tesla as a function of lead length.
Materials and methods: Five custom lead lengths, 20-60 cm, were exposed to a uniform magnitude and phase radiofrequency electric field to examine the effect of lead length on pacemaker lead tip heating for pacemaker-attached and abandoned pacemaker leads.
Results: Abandoned and pacemaker-attached leads show resonant heating behavior and maximum heating occurs at different lead lengths due to the differences in termination conditions. For clinical lead lengths (40-60 cm) abandoned leads exhibited greater lead tip heating compared with pacemaker-attached leads.
Conclusion: Current recommendations for MRI pacemaker safety should highlight the possible increased risk for patients with abandoned leads as compared to pacemaker-attached leads.
Copyright © 2011 Wiley-Liss, Inc.
Similar articles
-
Impact of capped and uncapped abandoned leads on the heating of an MR-conditional pacemaker implant.Magn Reson Med. 2015 Jan;73(1):390-400. doi: 10.1002/mrm.25106. Epub 2014 Jan 16. Magn Reson Med. 2015. PMID: 24436030
-
Numerical model for estimating RF-induced heating on a pacemaker implant during MRI: experimental validation.IEEE Trans Biomed Eng. 2010 Aug;57(8):2045-52. doi: 10.1109/TBME.2010.2043102. Epub 2010 Feb 18. IEEE Trans Biomed Eng. 2010. PMID: 20176532
-
In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometry.J Magn Reson Imaging. 2008 Oct;28(4):879-86. doi: 10.1002/jmri.21536. J Magn Reson Imaging. 2008. PMID: 18821629
-
Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.Laryngoscope. 2006 Aug;116(8):1321-33. doi: 10.1097/01.mlg.0000230479.39551.4a. Laryngoscope. 2006. PMID: 16885731 Review.
-
MRI site planning adventures.J Am Coll Radiol. 2004 Nov;1(11):871-3. doi: 10.1016/j.jacr.2004.06.013. J Am Coll Radiol. 2004. PMID: 17411721 Review. No abstract available.
Cited by
-
Feasibility of MRI in patients with non-Pacemaker/Defibrillator metallic devices and abandoned leads.J Biomed Sci Eng. 2021 Mar;14(3):83-93. doi: 10.4236/jbise.2021.143009. Epub 2021 Mar 9. J Biomed Sci Eng. 2021. PMID: 33777288 Free PMC article.
-
Magnetic resonance imaging of patients with epicardial leads: in vitro evaluation of temperature changes at the lead tip.J Interv Card Electrophysiol. 2019 Dec;56(3):321-326. doi: 10.1007/s10840-019-00627-7. Epub 2019 Nov 12. J Interv Card Electrophysiol. 2019. PMID: 31713219
-
Offline impedance measurements for detection and mitigation of dangerous implant interactions: an RF safety prescreen.Magn Reson Med. 2015 Mar;73(3):1328-39. doi: 10.1002/mrm.25202. Epub 2014 Mar 12. Magn Reson Med. 2015. PMID: 24623586 Free PMC article.
-
Temporary device malfunction of an MR conditional cardiac resynchronization defibrillator when undergoing MRI without appropriate re-programming: a case report.Eur Heart J Case Rep. 2021 Sep 21;5(9):ytab198. doi: 10.1093/ehjcr/ytab198. eCollection 2021 Sep. Eur Heart J Case Rep. 2021. PMID: 34557630 Free PMC article.
-
Safety of implanted cardiac devices in an MRI environment.Curr Cardiol Rep. 2015 Jul;17(7):605. doi: 10.1007/s11886-015-0605-9. Curr Cardiol Rep. 2015. PMID: 26026996 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical