Primary cup fixation with different designs of trapeziometacarpal total joint trapezium components: a radiostereometric analysis in a pig bone model
Abstract
High loosening rates of the trapezium components of trapeziometacarpal total joint prostheses have been reported. The purpose of this study was to compare the primary press fit fixation of two different, uncemented cup designs (MOTEC and Elektra) with the primary fixation of a cemented polyethylene cup (DLC) in a pig bone model. We did a push out test to measure the maximal load strength of the implants and a low-pressure cyclic loading test combined with radiostereometric analysis to measure the micromotion of the implants. There was no significant difference in fixation strength between the two uncemented screw cup designs or between the two uncemented screw cups and the cemented polyethylene cup. However, we found that threading of the bone before insertion of the Electra screw cup weakens the primary fixation strength of the implant. The results indicate that focus should be on the insertion technique as well as on the cup design of uncemented trapezium cup implants. Further studies of trapezium implant migration in a clinical setting are needed.
Similar articles
-
Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up.Acta Orthop. 2013 Feb;84(1):98-105. doi: 10.3109/17453674.2013.765625. Epub 2013 Jan 23. Acta Orthop. 2013. PMID: 23343372 Free PMC article. Clinical Trial.
-
Fixation of trapezial implants in a trapeziometacarpal total joint prosthesis tested in a model of porcine bone.J Plast Surg Hand Surg. 2011 Dec;45(6):263-6. doi: 10.3109/2000656X.2011.634564. J Plast Surg Hand Surg. 2011. PMID: 22250717
-
Trapeziometacarpal joint implants can be evaluated by roentgen stereophotogrammetric analysis.J Hand Surg Eur Vol. 2010 Jul;35(6):480-5. doi: 10.1177/1753193409355627. Epub 2010 Feb 24. J Hand Surg Eur Vol. 2010. PMID: 20181771
-
New polymer materials in total hip arthroplasty. Evaluation with radiostereometry, bone densitometry, radiography and clinical parameters.Acta Orthop Suppl. 2005 Feb;76(315):3-82. Acta Orthop Suppl. 2005. PMID: 15790289 Review.
-
Adjuvant therapies of bone graft around non-cemented experimental orthopedic implants stereological methods and experiments in dogs.Acta Orthop Suppl. 2008 Aug;79(330):1-43. Acta Orthop Suppl. 2008. PMID: 19065776 Review.
Cited by
-
The accuracy and precision of radiostereometric analysis in upper limb arthroplasty.Acta Orthop. 2017 Jun;88(3):320-325. doi: 10.1080/17453674.2017.1291872. Epub 2017 Mar 2. Acta Orthop. 2017. PMID: 28464752 Free PMC article.
-
Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up.Acta Orthop. 2013 Feb;84(1):98-105. doi: 10.3109/17453674.2013.765625. Epub 2013 Jan 23. Acta Orthop. 2013. PMID: 23343372 Free PMC article. Clinical Trial.
-
Trapezium Bone Density-A Comparison of Measurements by DXA and CT.J Funct Biomater. 2018 Jan 18;9(1):9. doi: 10.3390/jfb9010009. J Funct Biomater. 2018. PMID: 29346300 Free PMC article.
-
Outcome of uncemented trapeziometacarpal prosthesis for treatment of thumb carpometacarpal joint arthritis.Indian J Orthop. 2014 Jul;48(4):394-8. doi: 10.4103/0019-5413.136270. Indian J Orthop. 2014. PMID: 25143644 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical